Yokogawa PC-Based MX100 User Manual
Page 112
4-24
IM MX100-01E
4.8 4-CH, Medium-Speed Strain Input Module
(MX112) Specifications
Style number:
S2
Number of inputs:
4
Input type:
Strain gauge or strain gauge type sensors (static strain)
Input method:
Floating balanced input, islated between channels (NDIS is non-
isolated)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
One-Gauge Method Conversion Value
Input
Measurement
Range Type
Rated
Measurement
Range
Measurement
Accuracy
Resolution Measurement
Accuracy
Resolution
Integration time: 16.67 ms
or more
Integration time: 1.67 ms
Strain
2000 µStrain
±2000.0 µStrain
±0.5% of range 0.1 µStrain ±2% of range 1 µStrain
*1
20000 µStrain
±20000 µStrain
±0.3% of range 1 µStrain
±1% of range 2 µStrain
*2
200000 µStrain ±200000 µStrain ±0.3% of range 10 µStrain
±1% of range 10 µStrain
*1 Display resolution is 0.1 µStrain
*2 Display resolution is 1 µStrain
AD resolution:
Equivalent to ±20000 FS display
However, excludes 1.67 ms integration time
AD integration time:
Measurement Interval
Integration Time
Filter
Rejected Noise and Notes
100 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
200 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply
frequency and set 16.67 or 20 ms
500 ms
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
1 s
100 ms
Rectangular
10 Hz and its integer multiples
2, 5,10, 20, 30, 60 s
200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In such
cases, set the measurement interval to 200 ms or higher.
Gauge connection method:
1-gauge (2 or 3 wire systems), opposing 2-gauge, adjacent 2- or
4-gauge
With clamp terminals, set on a channel basis with switches
Applicable gauge resistance:
100-1000 Ω
Built in resistance of 120 Ω for -B12, and 350 Ω for -B35.
Bridge voltage:
Fixed at 2 VDC, accuracy ±5%, compensated with internal Cal
Applicable gauge factor: Fixed at 2.0. Gauge factor can be compensated with the scaling
function.
Balance adjustment: Automatic, digital calculation methods
Balance adjustment range: ±10000 µstrain (1-gauge method conversion)
Balance adjustment accuracy: The measurement accuracy or less:
Resistance accuracy for bridge: ±0.01% ±5ppm/°C
Input resistance:
1 MΩ or more
Allowable wiring resistance: 100 Ω or less
Effect of wiring resistance: NDIS 50 ppm of rdg/Ω (when using remote sensing wire).
Does not compensate for clamp wiring resistance. Depends on
the gauge resistance.