4.8 4-CH, Medium-Speed Strain Input Module

(MX112) Specifications

Style number:

S2

Number of inputs:

4

Input type:

Strain gauge or strain gauge type sensors (static strain)

Input method:

Floating balanced input, islated between channels (NDIS is non-
isolated)

Measurement range and accuracy:

The accuracy applies to standard operating conditions:

Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.

One-Gauge Method Conversion Value

Input

Measurement

Range Type

Rated

Measurement

Range

Measurement

Accuracy

Resolution Measurement

Accuracy

Resolution

Integration time: 16.67 ms

or more

Integration time: 1.67 ms

Strain

2000 µStrain

±2000.0 µStrain

±0.5% of range 0.1 µStrain ±2% of range 1 µStrain

*1

20000 µStrain

±20000 µStrain

±0.3% of range 1 µStrain

±1% of range 2 µStrain

*2

200000 µStrain ±200000 µStrain ±0.3% of range 10 µStrain

±1% of range 10 µStrain

*1 Display resolution is 0.1 µStrain
*2 Display resolution is 1 µStrain

AD resolution:

Equivalent to ±20000 FS display

However, excludes 1.67 ms integration time

AD integration time:

Measurement Interval

Integration Time

Filter

Rejected Noise and Notes

100 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

200 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply

frequency and set 16.67 or 20 ms

500 ms

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

1 s

100 ms

Rectangular

10 Hz and its integer multiples

2, 5,10, 20, 30, 60 s

200 ms

Cos

Fc = 5-Hz low-pass filter

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In such

cases, set the measurement interval to 200 ms or higher.

Gauge connection method:

1-gauge (2 or 3 wire systems), opposing 2-gauge, adjacent 2- or
4-gauge

With clamp terminals, set on a channel basis with switches

Applicable gauge resistance:

100-1000 Ω

Built in resistance of 120 Ω for -B12, and 350 Ω for -B35.

Bridge voltage:

Fixed at 2 VDC, accuracy ±5%, compensated with internal Cal

Applicable gauge factor: Fixed at 2.0. Gauge factor can be compensated with the scaling

function.

Balance adjustment: Automatic, digital calculation methods
Balance adjustment range: ±10000 µstrain (1-gauge method conversion)
Balance adjustment accuracy: The measurement accuracy or less:
Resistance accuracy for bridge: ±0.01% ±5ppm/°C
Input resistance:

1 MΩ or more

Allowable wiring resistance: 100 Ω or less
Effect of wiring resistance: NDIS 50 ppm of rdg/Ω (when using remote sensing wire).

Does not compensate for clamp wiring resistance. Depends on
the gauge resistance.