4-19
IM MX100-01E
Specification
3
2
1
4
5
Index
Effects of Operating Conditions
The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:
At least 30 minutes after power-up
Effects of ambient temperature:
The effect received by changes in 10°C increments of the ambient
temperature is
±(0.05% of rdg + 0.05% of range) or less
Effects of power fluctuation:
Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V
Effects of magnetic field:
The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is
±(0.1% of rdg + 10 digits) or less
Effects from the signal source resistance:
Effects on DC voltage and thermocouple from a fluctuation of 1
kΩ in the signal source resistance
DC voltage: 1 V range or less
±10 µV or less
2 V range or higher
±0.15% of rdg or less
Thermocouple: ±10 µV or less
Effect of position:
Horizontal with the feet at the bottom is the rule.
Effects of vibration:
The fluctuation that results by applying a sinusoidal vibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is ±(0.1% of rdg + 1 digit) or less
External Dimensions
Unit: mm
131
174
150
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.
4.6 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30) Specifications