Index – Yokogawa PC-Based MX100 User Manual
Page 105
4-17
IM MX100-01E
Specification
3
2
1
4
5
Index
The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).
Input
Measurement
Range Type
Rated Measurement
Range
Measurement Accuracy
Highest
Resolution
(1 Digit)
Integration Time: 16.67 ms
or More
Integration Time: 1.67 ms
DC voltage
60 mV (high res.)
0.000 to 60.000 mV
±(0.05% of rdg + 20 digits)
±(0.1% of rdg + 100 digits)
1 µV
1 V
–1.0000 to 1.0000 V
±(0.05% of rdg + 2 digits)
±(0.1% of rdg + 10 digits)
100 µV
6 V (high res.)
0.0000 to 6.0000 V
±(0.05% of rdg + 20 digits)
±(0.1% of rdg + 100 digits)
100 µV
Thermocouple
(excludes RJC accuracy
, when
burnout is OFF)
PLATINEL
0.0 to 1400.0°C
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 4°C)
0.1°C
PR40-20
0.0 to 1900.0°C
±(0.05% of rdg + 2.5°C)
Except 300 to 700°C: ±6°C
less than 300°C: not guaranteed
±(0.1% of rdg + 12°C)
Except 300 to 700°C: ±25°C
less than 300°C: not guaranteed
NiNiMo
0.0 to 1310.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 2.7°C)
WRe3-25
0.0 to 2400.0°C
±(0.05% of rdg + 2°C)
Except 0 to 200°C: ±2.5°C
2000°C or higher:
±(0.05% of rdg + 4°C)
±(0.1% of rdg + 7°C)
Except 0 to 200°C: ±12°C
2000°C or higher:
±(0.1% of rdg + 11°C)
W/WRe26
0.0 to 2400.0°C
±(0.05% of rdg + 2°C)
Except 100 to 300°C: ±4°C
less than 100°C: not guaranteed
±(0.1% of rdg + 8.5°C)
Except 100 to 300°C: ±12°C
less than 100°C: not guaranteed
N (AWG14)
0.0 to 1300.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 3.5°C)
XK GOST
–200.0 to 600.0°C
±(0.05% of rdg + 0.5°C)
Except –200 to 0°C:
±(0.2% of rdg + 0.7°C)
±(0.1% of rdg + 2.5°C)
Except –200 to 0°C:
±(1% of rdg + 2.5°C)
Measurement interval, integration time, and filter:
Measurement Interval Integration Time
Filter
Rejected Noise and Notes
500 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
1 s
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply
frequency and set 16.67 or 20 ms
2 s
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
5,10, 20, 30, 60 s
100 ms
Rectangular
10 Hz and its integer multiples
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially
with temperature measurement. In such cases, set the measurement interval to 1 s or higher, or use
the 4-CH, High-Speed Universal Input Module or the 10-CH, Medium-Speed Universal Input Module.
Reference junction compensation:
Switch between external and internal on each channel, remote
RJC function available
Reference junction compensation accuracy:
When measuring temperature greater than or equal to 0°C and
when the temperature of the input terminal is balanced
Type R, S, W: ±1°C
Type K, J, E, T, N, L, U, XK GOST: ±0.5°C
Type N(AWG14), PLATINEL, NiNiMo, WRe3-25, W/WRe26: ±1°C
Note: The internal reference junction compensation is fixed to 0°C
for type B and PR40-20
Maximum input voltage:
DC voltage at 1-V range or less, thermocouple, and DI (contact):
±10 VDC (continuous)
Other measurement ranges: ±120 VDC (continuous)
Normal-mode voltage:
DC voltage, thermocouple, DI (LEVEL): 1.2 times the range rating
or less (50/60 Hz, peak value including the signal component)
Normal-mode rejection ratio:
For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)
For integration time of 1.67 ms: 50/60 Hz is not rejected
4.6 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30) Specifications