beautypg.com

6 connecting the ethernet cable, Connection procedure, Connecting the ethernet cable -22 – Yokogawa PC-Based MX100 User Manual

Page 67: Connection procedure -22

background image

2-22

IM MX100-01E

2.6 Connecting the Ethernet Cable

Connection Procedure

Connectors

Connect the Ethernet cable to the Ethernet port (10BASE-T/100BASE-TX) of the main
module. Use a UTP cable (category 5 or better) or an STP cable for the Ethernet cable.

RJ-45 modular jack

Ethernet port

Ethernet cable

Connection to the PC

Make the connection via a hub. For one-to-one connection with a PC, connect as shown
in the figure below. Likewise, up to twenty MX100s can be connected to a single PC.
However, to connect two or more MX100s, MXLOGGER (sold separately) must be used
instead of the MX100 Standard Software that comes with the instrument.

MX100

Hub

MX100

Ethernet port

PC

LAN cable

(straight cable)

Note

The NIC on the PC should support 100BASE-TX (recommended) or 10BASE-T.

When connected to an external network, the communications within the network other than

those related to the MX100 may hinder the measurement operations on the MX100.

The MX100 can be connected to an auto-negotiating hub (between 10BASE-T and

100BASE-TX), but connection may fail on some hubs due to the hub not recognizing the
signal output from the MX100. If this happens, you can turn only dip switch 6 of the main
module OFF to fix the data rate to 10 Mbps. Operate the dip switch only when the power is
turned OFF.