2.3 Attaching the Modules

WARNING

To prevent electric shock and instrument breakdown, do not connect the power
supply to the main module when attaching modules.

Attachment Procedure

1.

Check that the power supply is not connected to the main module.

2.

Align the connector on the rear panel of the module to the module connector on

the base plate, and insert the connector.

Attach the main module to the right-most position on the base plate.
When the connectors are correctly connected, the guide pin on the rear panel of the module
is inserted into the guide hole on the base plate. In addition, the module is secured to the
base plate with the latch lever locking in place at the bottom section of the base plate.

3.

Fasten the top of the module with the attachment screws (M3).

To remove the module, loosen the attachment screw, pull down on the
latch lever on the rear panel of the module, and pull the module straight
from the base plate.

Module connector

Base plate connector

cover*

(accessory sold

separately)

Screw holes

Guide holes

Main module

Input/Output module

Base plate

Base plate

Latch lever

(pull down to remove the

module from the base plate)

Guide pins

Module attachment screw

* Can be attached by

hooking the top of the

cover to the top edge of

the base plate and

pressing down on the

bottom part.