Index – Yokogawa PC-Based MX100 User Manual
Page 70
2-25
IM MX100-01E
Installation and W
iring
3
2
1
4
5
Index
On the MX100, the integral time is determined by the measurement interval setting as
shown in the table below.
4-CH, High-Speed Universal Input Module
Measurement
Interval
Integration
Time
Filter
Rejected Noise and Notes
10 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
50 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
100, 200 ms
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
500 ms
100 ms
Rectangular
10 Hz and its integer multiples
1, 2, 5,10, 20,
30, 60 s
200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 50
ms or higher.
10-CH, Middle-Speed Universal Input Module/6-CH, Medium-Speed Four-Wire RTD
Resistance Input Module
Measurement
Interval
Integration
Time
Filter
Rejected Noise and Notes
100, 200 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
500 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
1 s
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
2 s
100 ms
Rectangular
10 Hz and its integer multiples
5, 10, 20, 30, 60 s 200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 500
ms or higher, or use the 4-CH, High-Speed Universal Input Module
30-CH, Middle-Speed DCV/TC/DI Input Module
Measurement
Interval
Integration
Time
Filter
Rejected Noise and Notes
500 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
1 s
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
2 s
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
5, 10, 20, 30, 60 s 100 ms
Rectangular
10 Hz and its integer multiples
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 1 s
or higher, or use the 4-CH, High-Speed Universal Input Module or the 10-CH, Medium-Speed
Universal Input Module.
4-CH, Medium-Speed Strain Input Module
Measurement
Interval
Integration
Time
Filter
Rejected Noise and Notes
100 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
200 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
500 ms
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
1 s
100 ms
Rectangular
10 Hz and its integer multiples
2, 5,10, 20,
30, 60 s
200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In
such cases, set the measurement interval to 200 ms or higher.
2.7 Measures against Noise on the MX100