4-33
IM MX100-01E
Specification
3
2
1
4
5
Index
4.13 10-CH, Medium-Speed Digital Output Module
(MX125) Specifications
Style number:
S1
Number of outputs:
10
Contact mode:
A contact (SPST)
Update interval:
Output per 100 ms (not synchronized to the measurement
interval)
Contact capacity:
250 VDC/0.1 A, 250 VAC/2 A, or 30 VDC/2A (resistance load)
Contact life*:
100,000 times at rated load (typical)
20,000,000 times at no load (typical)
* The contact life varies depending on the load conditions and the
environment in which it is used.
Insulation resistance: 20 MΩ or more (500 VDC) between output terminal and earth
20 MΩ or more (500 VDC) between the output terminals
Withstand voltage:
2300 VAC (50/60 Hz) for one minute between the output terminal
and earth
2300 VAC (50/60 Hz) for one minute between output terminals
Maximum common-mode voltage:
250 VACrms (50/60 Hz)
Power consumption: Approximately 2 W (when all relays are turned ON)
Terminal type:
Clamp, attached and removed in units of 5 channels
Applicable wire size: 0.08 to 2.5 mm
2
(AWG28 to 12)
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal
cover)
Weight:
Approximately 0.5 kg
External Dimensions
Unit: mm
57
13
1
150.8
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.