Dual Save Functions (/DS Option)
3
2
1
4
5
Index
Dual Save Function (When Using an Application Based on the MX100/
DARWIN API
• Application Example 1
Backup
File No.=0
No.=1
Info
Backup
ON
Idle
Start
Stop
Backup
OFF
Measure
Stop Start
No.=2
Backup
Measure
GetData
Data Handling
(60 s)
Mode
Data
CF Write
CF File
Save on CF
MX100/
Daewin
API
application
Start
Start
New file
Stop
Stop
Data handling on the PC after Using GetData (Display, Computation, Thinning,
Saving, Etc.)
• Application Example 2
FileNo.=3
Info
Backup
ON
Idle
Start
Stop
Backup
OFF
Measure
Stop
Start
No.=4
Start
Measure
Idle
Idle
GetData
Data Handling
Mode
Data
CF Write
CF File
MX100/
Daewin
API
application
Backup
Backup
Start
Start
Stop
Stop
Save on CF
(60 s)
You can set recording status settings before and after measurement starts or stops
(Backup ON/OFF). Saving of data to the CF card can be linked with the Start/Stop
function.
Note
The file number of a data file saved to the CF card using an MX100/DARWIN API application is
the consecutive number assigned during use of the API application.
Output Operation of the Dual Save Function
If the Dual Save function is used, output from digital, analog, and PWM output modules
may experience delay when data is being saved to the CF card.
When stopping recording in particular, a delay of several seconds may occur due to the
save stop operation.