Allowable input voltage: ±10 VDC (between H-L) continuous
Allowable common-mode voltage:

Channel-to-channel: 30 VAC rms

Between input and earth:250 VAC rms (-B12,-B35) 30 VACrms(-NDI)

However, NDIS connector shell is connected to earth potential

CMRR

*

:

For integration time of 16.67 ms or more: 120 dB or more

For integration time of 1.67 ms or more: 80 dB or more

(voltage conversion value at 50/60 Hz ±0.1%, bridge voltage of 2 V)

NMRR:

For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)

For integration time of 1.67 ms: 50/60 Hz is not rejected.

(voltage conversion value with bridge voltage of 2 V)

Insulation resistance

*

: Between input and earth:

20 MΩ or more at 500 VDC

Withstand voltage

*

:

Between input and earth:

2300 VAC for one minute

Channel-to-channel:

30 VACrms or less

Power consumption: Approximately 3 W (one module)
Weight:

Approximately 0.5 kg

External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal cover)
Terminal type:

-B12, -B35: Clamp, terminal board is detachable

-NDI:

NDIS

Applicable wire size: 0.14 to 1.5 mm

2

(AWG26 to 16)(excluding -NDI)

* Not applicable in the case of the NDIS terminal.

Effects of Operating Conditions

The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:

At least 30 minutes after power-up

Effects of ambient temperature:

The effect received by changes in 10°C increments of the ambient
temperature is:

±(0.1% of range) or less

Effects of power fluctuation:

Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V

Effects of magnetic field:

The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is ±2% or range or less

Effect of position:

Horizontal with the feet at the bottom is the rule.

External Dimensions

Unit: mm

57

150.8

133.8

57

4.8

-B12, -B35

-NDI

13

1

131

If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3

mm.

4.8 4-CH, Medium-Speed Strain Input Module (MX112) Specifications