Yokogawa PC-Based MX100 User Manual
Page 104
4-16
IM MX100-01E
4.6 30-CH, Medium-Speed DCV/TC/DI Input Module
(MX110-VTD-L30) Specifications
Style number:
S3
Type of measurement: DC voltage, thermocouple, DI (LEVEL, non-voltage contact)
Number of inputs:
30
Module width:
3 modules wide
Input method:
Floating unbalanced input, insulation between channels
A/D resolution:
16 bits (±20000/±6000/0 to 60000)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
Input
Measurement
Range Type
Rated Measurement
Range
Measurement Accuracy
Highest
Resolution
(1 Digit)
Integration Time: 16.67 ms
or More
Integration Time: 1.67 ms
DC voltage
20 mV
–20.000 to 20.000 mV ±(0.05% of rdg + 5 digits)
±(0.1% of rdg + 25 digits)
1 µV
60 mV
–60.00 to 60.00 mV
±(0.05% of rdg + 2 digits)
±(0.1% of rdg + 10 digits)
10 µV
200 mV
–200.00 to 200.00 mV
2 V
–2.0000 to 2.0000 V
±(0.05% of rdg + 5 digits)
100 µV
6 V
–6.000 to 6.000 V
±(0.05% of rdg + 2 digits)
1 mV
20 V
–20.000 to 20.000 V
100 V
–100.00 to 100.00 V
10 mV
Thermocouple
(excludes RJC accuracy
, when burnout
is OFF)
R
*1
0.0 to 1760.0°C
±(0.05% of rdg + 1°C)
Except 0 to 100°C: ±3.7°C, 100
to 300°C: ±1.5°C for R and S;
400 to 600°C: ±2°C, less than
400°C: not guaranteed for B
±(0.1% of rdg + 4°C)
Except 0 to 100°C: ±10°C, 100
to 300°C: ±5°C for R and S;
400 to 600°C: ±7°C, less than
400°C: not guaranteed for B
0.1°C
S
*1
B
*1
0.0 to 1820.0°C
K
*1
–200.0 to 1370.0°C
±(0.05% of rdg + 0.7°C)
Except –200°C to –100°C:
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 3.5°C)
Except –200°C to –100°C:
±(0.1% of rdg + 6°C)
E
*1
–200.0 to 800.0°C
±(0.05% of rdg + 0.5°C)
Except –200°C to –100°C:
±(0.05% of rdg + 0.7°C) for J
and L
±(0.1% of rdg + 2.5°C)
Except –200°C to –100°C:
±(0.1% of rdg + 5°C)
J
*1
–200.0 to 1100.0°C
T
*1
–200.0 to 400.0°C
L
*2
–200.0 to 900.0°C
U
–200.0 to 400.0°C
N
*3
0.0 to 1300.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 3.5°C)
W
*4
0.0 to 2315.0°C
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 7°C)
KPvsAu7Fe
0.0 to 300.0K
±(0.05% of rdg + 0.7K)
±(0.1% of rdg + 3.5K)
0.1K
DI
Level
Vth = 2.4 V
Threshold level accuracy: ±0.1 V
Contact input
ON for 1 kΩ or less, OFF for 100 kΩ or more (parallel capacitance: 0.01 µF or less)
*5
*1 R, S, B, K, E, J, T: ANSI, IEC 584, DIN IEC 584, JIS C 1602-1995
*2 L: Fe-CuNi, DIN43710/U: Cu-CuNi, DIN 43710
*3 N: Nicrosil-Nisil, IEC 584, DIN IEC 584
*4 W: W-5% RE/W-26% Re (Hoskins Mfg Co)
*5 Measured using a measurement current of approximately 10 µA at 200-mV range. Threshold level is approximately 0.1 V.