Yokogawa PC-Based MX100 User Manual
Page 96
4-8
IM MX100-01E
4.4 4-CH, High-Speed Universal Input Module
(MX110-UNV-H04) Specifications
Style number:
S1
Type of measurement: DC voltage, thermocouple, 3-wire RTD, DI (LEVEL, non-voltage contact)
Number of inputs:
4
Input method:
Floating unbalanced input, insulation between channels
A/D resolution:
16 bits (±20000/±6000/0 to 60000)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
Input
Measurement
Range Type
Rated Measurement
Range
Measurement Accuracy
Highest
Resolution
(1 Digit)
Integration Time: 16.67 ms
or More
Integration Time: 1.67 ms
DC voltage
20 mV
–20.000 to 20.000 mV ±(0.05% of rdg + 5 digits)
±(0.1% of rdg + 25 digits)
1 µV
60 mV
–60.00 to 60.00 mV
±(0.05% of rdg + 2 digits)
±(0.1% of rdg + 10 digits)
10 µV
200 mV
–200.00 to 200.00 mV
2 V
–2.0000 to 2.0000 V
±(0.05% of rdg + 5 digits)
100 µV
6 V
–6.000 to 6.000 V
±(0.05% of rdg + 2 digits)
1 mV
20 V
–20.000 to 20.000 V
100 V
–100.00 to 100.00 V
10 mV
Thermocouple
(excludes RJC accuracy
, when burnout
is OFF)
R
*1
0.0 to 1760.0°C
±(0.05% of rdg + 1°C)
Except 0 to 100°C: ±3.7°C, 100
to 300°C: ±1.5°C for R and S;
400 to 600°C: ±2°C, less than
400°C: not guaranteed for B
±(0.1% of rdg + 4°C)
Except 0 to 100°C: ±10°C, 100
to 300°C: ±5°C for R and S;
400 to 600°C: ±7°C, less than
400°C: not guaranteed for B
0.1°C
S
*1
B
*1
0.0 to 1820.0°C
K
*1
–200.0 to 1370.0°C
±(0.05% of rdg + 0.7°C)
Except –200°C to –100°C:
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 3.5°C)
Except –200°C to –100°C:
±(0.1% of rdg + 6°C)
E
*1
–200.0 to 800.0°C
±(0.05% of rdg + 0.5°C)
Except –200°C to –100°C:
±(0.05% of rdg + 0.7°C) for J
and L
±(0.1% of rdg + 2.5°C)
Except –200°C to –100°C:
±(0.1% of rdg + 5°C)
J
*1
–200.0 to 1100.0°C
T
*1
–200.0 to 400.0°C
L
*2
–200.0 to 900.0°C
U
–200.0 to 400.0°C
N
*3
0.0 to 1300.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 3.5°C)
W
*4
0.0 to 2315.0°C
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 7°C)
KPvsAu7Fe
0.0 to 300.0K
±(0.05% of rdg + 0.7K)
±(0.1% of rdg + 3.5K)
0.1K
R
TD
(Measurement current: 1 mA)
Pt100
*5
–200.0 to 600.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.1°C
JPt100
*5
–200.0 to 550.0°C
Pt100 (high res.)
–140.00 to 150.00°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.01°C
JPt100 (high res.)
–140.00 to 150.00°C
Ni100 SAMA
*6
–200.0 to 250.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.1°C
Ni100 DIN
*6
–60.0 to 180.0°C
Ni120
*7
–70.0 to 200.0°C
R
TD
(Measurement current:
2 mA)
Pt100
*5
–200.0 to 250.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.1°C
JPt100
*5
–200.0 to 250.0°C
Pt100 (high res.)
–140.00 to 150.00°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.01°C
JPt100 (high res.)
–140.00 to 150.00°C
Pt50
*5
–200.0 to 550.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
0.1°C
Cu10 GE
*8
–200.0 to 300.0°C
±(0.05% of rdg + 0.7°C)
±(0.2% of rdg + 2.5°C)
Cu10 L&N
*8
–200.0 to 300.0°C
Cu10 WEED
*8
–200.0 to 300.0°C
Cu10 BAILEY
*8
–200.0 to 300.0°C
J263B
0.0 to 300.0K
±(0.05% of rdg + 0.3K)
±(0.1% of rdg + 1.5K)
0.1K
DI
Level
Vth = 2.4 V
Threshold level accuracy: ±0.1 V
Contact input
ON for 100 Ω or less, OFF for 10 kΩ or more
*9
*1 R, S, B, K, E, J, T: ANSI, IEC 584, DIN IEC 584, JIS C 1602-1995
*2 L: Fe-CuNi, DIN43710/U: Cu-CuNi, DIN 43710
*3 N: Nicrosil-Nisil, IEC 584, DIN IEC 584
*4 W: W-5% RE/W-26% Re (Hoskins Mfg Co)
*5 Pt50: JIS C 1604-1989, JIS C 1606-1989/Pt100: JIS C 1604-1997, IEC 751, DIN IEC 751/JPt100: JIS C 1604-1989, JIS C 1606-1989
*6 SAMA/DIN
*7 McGRAW EDISON COMPANY
*8 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/
Cu10 BAILEY: –20.0 to 250.0°C
*9 Measured using a measurement current of approximately 1 mA at 1-V range. Threshold level is approximately 0.8 V.