4.4 4-CH, High-Speed Universal Input Module

(MX110-UNV-H04) Specifications

Style number:

S1

Type of measurement: DC voltage, thermocouple, 3-wire RTD, DI (LEVEL, non-voltage contact)
Number of inputs:

4

Input method:

Floating unbalanced input, insulation between channels

A/D resolution:

16 bits (±20000/±6000/0 to 60000)

Measurement range and accuracy:

The accuracy applies to standard operating conditions:

Ambient temp: 23 ±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.

Input

Measurement

Range Type

Rated Measurement

Range

Measurement Accuracy

Highest

Resolution

(1 Digit)

Integration Time: 16.67 ms

or More

Integration Time: 1.67 ms

DC voltage

20 mV

–20.000 to 20.000 mV ±(0.05% of rdg + 5 digits)

±(0.1% of rdg + 25 digits)

1 µV

60 mV

–60.00 to 60.00 mV

±(0.05% of rdg + 2 digits)

±(0.1% of rdg + 10 digits)

10 µV

200 mV

–200.00 to 200.00 mV

2 V

–2.0000 to 2.0000 V

±(0.05% of rdg + 5 digits)

100 µV

6 V

–6.000 to 6.000 V

±(0.05% of rdg + 2 digits)

1 mV

20 V

–20.000 to 20.000 V

100 V

–100.00 to 100.00 V

10 mV

Thermocouple

(excludes RJC accuracy

, when burnout

is OFF)

R

*1

0.0 to 1760.0°C

±(0.05% of rdg + 1°C)

Except 0 to 100°C: ±3.7°C, 100

to 300°C: ±1.5°C for R and S;

400 to 600°C: ±2°C, less than

400°C: not guaranteed for B

±(0.1% of rdg + 4°C)

Except 0 to 100°C: ±10°C, 100

to 300°C: ±5°C for R and S;

400 to 600°C: ±7°C, less than

400°C: not guaranteed for B

0.1°C

S

*1

B

*1

0.0 to 1820.0°C

K

*1

–200.0 to 1370.0°C

±(0.05% of rdg + 0.7°C)

Except –200°C to –100°C:

±(0.05% of rdg + 1°C)

±(0.1% of rdg + 3.5°C)

Except –200°C to –100°C:

±(0.1% of rdg + 6°C)

E

*1

–200.0 to 800.0°C

±(0.05% of rdg + 0.5°C)

Except –200°C to –100°C:

±(0.05% of rdg + 0.7°C) for J

and L

±(0.1% of rdg + 2.5°C)

Except –200°C to –100°C:

±(0.1% of rdg + 5°C)

J

*1

–200.0 to 1100.0°C

T

*1

–200.0 to 400.0°C

L

*2

–200.0 to 900.0°C

U

–200.0 to 400.0°C

N

*3

0.0 to 1300.0°C

±(0.05% of rdg + 0.7°C)

±(0.1% of rdg + 3.5°C)

W

*4

0.0 to 2315.0°C

±(0.05% of rdg + 1°C)

±(0.1% of rdg + 7°C)

KPvsAu7Fe

0.0 to 300.0K

±(0.05% of rdg + 0.7K)

±(0.1% of rdg + 3.5K)

0.1K

R

TD

(Measurement current: 1 mA)

Pt100

*5

–200.0 to 600.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

JPt100

*5

–200.0 to 550.0°C

Pt100 (high res.)

–140.00 to 150.00°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.01°C

JPt100 (high res.)

–140.00 to 150.00°C

Ni100 SAMA

*6

–200.0 to 250.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

Ni100 DIN

*6

–60.0 to 180.0°C

Ni120

*7

–70.0 to 200.0°C

R

TD

(Measurement current:

2 mA)

Pt100

*5

–200.0 to 250.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

JPt100

*5

–200.0 to 250.0°C

Pt100 (high res.)

–140.00 to 150.00°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.01°C

JPt100 (high res.)

–140.00 to 150.00°C

Pt50

*5

–200.0 to 550.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

Cu10 GE

*8

–200.0 to 300.0°C

±(0.05% of rdg + 0.7°C)

±(0.2% of rdg + 2.5°C)

Cu10 L&N

*8

–200.0 to 300.0°C

Cu10 WEED

*8

–200.0 to 300.0°C

Cu10 BAILEY

*8

–200.0 to 300.0°C

J263B

0.0 to 300.0K

±(0.05% of rdg + 0.3K)

±(0.1% of rdg + 1.5K)

0.1K

DI

Level

Vth = 2.4 V

Threshold level accuracy: ±0.1 V

Contact input

ON for 100 Ω or less, OFF for 10 kΩ or more

*9

*1 R, S, B, K, E, J, T: ANSI, IEC 584, DIN IEC 584, JIS C 1602-1995

*2 L: Fe-CuNi, DIN43710/U: Cu-CuNi, DIN 43710
*3 N: Nicrosil-Nisil, IEC 584, DIN IEC 584
*4 W: W-5% RE/W-26% Re (Hoskins Mfg Co)

*5 Pt50: JIS C 1604-1989, JIS C 1606-1989/Pt100: JIS C 1604-1997, IEC 751, DIN IEC 751/JPt100: JIS C 1604-1989, JIS C 1606-1989

*6 SAMA/DIN

*7 McGRAW EDISON COMPANY

*8 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/

Cu10 BAILEY: –20.0 to 250.0°C

*9 Measured using a measurement current of approximately 1 mA at 1-V range. Threshold level is approximately 0.8 V.