Yokogawa PC-Based MX100 User Manual
Page 76
3-4
IM MX100-01E
The MX100 can be detected using the Search button, but connection fails.
Probable Cause
Corrective Action
Ref. Page
The IP address is set to the default value.
The default value cannot be used to make the
connection.
Set the correct IP address.
*
There is a problem in the network
configuration.
Check that the IP address, subnet mask, and default gateway
settings of the PC and the MX100 settings are correct.
*
The PC software, main module style number,
and release number rule is not upheld.
Check the PC software release number and main module style
number before upgrading the version.
[PC software release no.] ≥ [main module style no.]
i
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
The Calibrator cannot connect to the MX100.
Probable Cause
Corrective Action
Ref. Page
Attempting to make multiple connections.
Another software program such as the MX100
Standard Software is already connected.
Close all other software programs such as the MX100 Standard
Software.
*
The main module and input/output module
style number rule is not upheld.
Check the release number of the PC software and the main
module style number, then upgrade the version.
[PC software release no.] ≥ [main module style no.] ≥ [input/output
module style no.]
i
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
The connected input/output module is not detected.
Probable Cause
Corrective Action
Ref. Page
Module connection or module startup error.
Attached the module while the power was ON.
Turn OFF the power. Detach the input/output module once and
attach it again.
2-4, 2-19
The PC software, main module, and input/
output module style and release number rule
is not upheld.
Check the PC software release number and main module style
number before upgrading the version.
[PC software release no.] ≥ [main module style no.] ≥ [input/output
module style no.]
i
Carried out an incorrect calibration.
Recalibrate.
*
The 10-CH, Pulse Input Module (MX114-
PLS-M10) is connected.
The 10-CH, Pulse Input Module cannot be used with the MX100.
Use this module with the MW100.
1-5
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
3.2 Troubleshooting