Troubleshooting and Maintenance

3

2

1

4

5

Index

• When calibrating the RTD range of the 4-CH, High-Speed Universal Input module

100Vpk MAX TO

250V MAX CH TO CH

600V MAX TO

100Vpk MAX TO

250V MAX CH TO CH

600V MAX TO

H
L
G

Resistance standard

Input terminal

Input terminal

Short

Make the resistance of

three lead wires equal.

When calibrating 0

When calibrating a range other than 0

* Carry out calibration for each input

terminal.

b

A

B

b

A

B

• When calibrating the DC voltage range of the 10-CH, Medium-Speed Universal Input

module

100Vpk MAX TO

120V MAX CH TO CH

600V MAX TO

Hi

Lo

Input terminal of CH2

Short the input terminal of CH1 (apply 0 V)

+

+

DC voltage/current standard

• When calibrating the RTD range of the 10-CH, Medium-Speed Universal Input module

b

A

B

b

A

B

100Vpk MAX TO

120V MAX CH TO CH

600V MAX TO

H
L
G

Short the input terminal of CH3 (connect 0 )

When calibrating the RTD (1 mA) 600 mV range,

connect 300 to the input terminal of CH5.*

Make the resistance of three lead wires equal.

* When calibrating the RTD (1 mA) 60 mV range, connect 60 to

the input terminal of CH6.

When calibrating the RTD (1 mA) 200 mV range, connect 200

to the input terminal of CH4.

Resistance standard

3.3 Calibration