4-6

IM MX100-01E

Display

LED:

Ethernet communication status (LINK, ACT)

Two-digit 7-segment LED: Status of the MX100 (unit number, operation status, and error

indication)

Communications

Interface:

Ethernet (10BASE-T/100BASE-TX)

Connector type:

RJ-45

Protocol:

TCP, IP, UDP, ARP, and ICMP

Transmission function: Transmits measured values and settings
Reception function:

Receives settings

TCP connection timeout: If there is no packet exchange for three minutes or more (not

configurable) when using TCP, a connection timeout occurs.
This mainly occurs when using the MX100/DARWIN API (sold
separately).

Other Functions

Time function:

Normally operates using the PC clock

Backup function available (records the times of the data using
time stamps of the main module when communication is
disconnected)

Internal clock accuracy: ±100 ppm
Dip switches:

Used to initialize settings, etc. (number of switches: 8)

Access stop switch (CF card control switch):

Closes the file to which data is being written on the CF card

Saves the log file

Memory backup

Settings and the clock are maintained by the internal lithium
battery, which has a lifetime of approximately 10 years (under
standard operating conditions)

General Specifications

Power consumption: Approximately 8 W
Insulation resistance: 20 MΩ or more at 500 VDC between the power supply terminal

and earth terminal

Withstand voltage:

1500 VAC (50/60 Hz) for one minute between the power supply
terminal and earth terminal

External dimensions: Approximately 92 (W) × 131 (H) × 137 (D) mm
Weight:

Approximately 0.85 kg

External Dimensions

92

136.7

13

1

Unit: mm

If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3

mm.