Index saving data to the cf card – Yokogawa PC-Based MX100 User Manual
Page 124
![background image](/manuals/360298/124/background.png)
5-3
IM MX100-01E
Dual Save Functions (/DS Option)
3
2
1
4
5
Index
Saving data to the CF Card
Transmission of measured data to the PC via communications and saving of measured
data to the CF card in the main module can be carried out simultaneously.
The standard function saves data to the CF card only if communication with the PC
is disconnected during recording by the PC software. With the
Dual Save function,
measured data is saved to the CF card during recording by the PC software.
This function is also available when programming applications with the MX100/DARWIN
API (use in combination with the “Backup”recording command and the “GetData”data
acquisition command).
Note
Under normal functionality, saving of data to the CF card is for backup when communications
are disconnected. The following shows the operation.
Backup When Communication Is Cut
Mode
Timeout
(60 s)
(60 s)
Connects (Monitor display)
Data backup
Command
Data
CF Write
CF File
PC Software
Communication
status
Recording
operation
Disconnects
No data displayed
Save on PC
Save on CF
Recording
starts
Recording
stops
Data acquired
and replenished
retroactively
Replenish
lost data
Comm. cut
Comm. restart
Backup
Backup
ON
Idle
Measure
Start
Stop
Backup
OFF
Measure
Idle
(GetData)
5.2 Explanation of Functions