4-23
IM MX100-01E
Specification
3
2
1
4
5
Index
Effects of Operating Conditions
The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:
At least 30 minutes after power-up
Effects of ambient temperature:
The effect received by changes in 10°C increments of the ambient
temperature is: ±(0.05% of rdg + 0.05% of range) or less
However, ±(0.2% of range + 1 digit) for Cu10 Ω
Effects of power fluctuation:
Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V
Effects of magnetic field:
The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is ±(0.1% of rdg + 10 digits) or less
Effects from the signal source resistance:
Effects on DC voltage from a fluctuation of 1 kΩ in the signal
source resistance
1 V range or less: ±10 µV or less
2 V range or higher: ±0.15% of rdg or less
RTD:
Fluctuations resulting from 10 Ω changes per wire:
1000 Ω and 100 Ω systems: ±0.1°C or less
Other than 1000 Ω and 100 Ω systems: ±1.0°C or less
Resistance: Fluctuations resulting from 10 Ω changes per wire: ±
1 digit or less
Effect of position:
Horizontal with the feet at the bottom is the rule.
Effects of vibration:
The fluctuation that results by applying a sinusoidal vibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is ±(0.1% of rdg + 1 digit) or less
External Dimensions
Unit: mm
57
150.8
13
1
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.