Specification

3

2

1

4

5

Index

Effects of Operating Conditions

The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:

At least 30 minutes after power-up

Effects of ambient temperature:

The effect received by changes in 10°C increments of the ambient
temperature is: ±(0.05% of rdg + 0.05% of range) or less

However, ±(0.2% of range + 1 digit) for Cu10 Ω

Effects of power fluctuation:

Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V

Effects of magnetic field:

The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is ±(0.1% of rdg + 10 digits) or less

Effects from the signal source resistance:

Effects on DC voltage from a fluctuation of 1 kΩ in the signal
source resistance

1 V range or less: ±10 µV or less

2 V range or higher: ±0.15% of rdg or less

RTD:

Fluctuations resulting from 10 Ω changes per wire:

1000 Ω and 100 Ω systems: ±0.1°C or less

Other than 1000 Ω and 100 Ω systems: ±1.0°C or less

Resistance: Fluctuations resulting from 10 Ω changes per wire: ±

1 digit or less

Effect of position:

Horizontal with the feet at the bottom is the rule.

Effects of vibration:

The fluctuation that results by applying a sinusoidal vibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s

2

is ±(0.1% of rdg + 1 digit) or less

External Dimensions

Unit: mm

57

150.8

13

1

If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3

mm.