background image

4-20

IM MX100-01E

4.7 6-CH, Medium-Speed Four-Wire RTD

Resistance Input Module (MX110-V4R-M06)

Specifications

Style number:

S2

Type of measurement: DC voltage, 4-wire RTD, 4-wire resistance, DI (LEVEL, non-

voltage contact)

Number of inputs:

6

Input method:

Floating unbalanced input, insulation between channels

A/D resolution:

16 bits (±20000/±6000/0 to 60000)

Measurement range and accuracy:

The accuracy applies to standard operating conditions:

Ambient temp: 23±2°C, ambient humidity: 55 ±10% RH, supply
voltage: 90 to 250 VAC, power frequency: 50/60 Hz ±1%, warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.

Input

Measurement

Range Type

Rated Measurement

Range

Measurement Accuracy

Highest

Resolution

(1 Digit)

Integration Time: 16.67 ms

or More

Integration Time: 1.67 ms

DC voltage

20 mV

–20.000 to 20.000 mV ±(0.05% of rdg + 5 digits)

±(0.1% of rdg + 25 digits)

1 µV

60 mV

–60.00 to 60.00 mV

±(0.05% of rdg + 2 digits)

±(0.1% of rdg + 10 digits)

10 µV

200 mV

–200.00 to 200.00 mV

2 V

–2.0000 to 2.0000 V

±(0.05% of rdg + 5 digits)

100 µV

6 V

–6.000 to 6.000 V

±(0.05% of rdg + 2 digits)

1 mV

20 V

–20.000 to 20.000 V

100 V

–100.00 to 100.00 V

10 mV

R

TD

*7

(Measurement current: 1 mA)

Pt100

*1

–200.0 to 600.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

JPt100

*1

–200.0 to 550.0°C

Pt100 (high res.)

–140.00 to 150.00°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.01°C

JPt100 (high res.)

–140.00 to 150.00°C

Ni100 SAMA

*2

–200.0 to 250.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

Ni100 DIN

*2

–60.0 to 180.0°C

Ni120

*3

–70.0 to 200.0°C

Pt50

*5

–200.0 to 550.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

Cu10 GE

*4

–200.0 to 300.0°C

±(0.1% of rdg + 2°C)

±(0.1% of rdg + 1.5°C)

Cu10 L&N

*4

–200.0 to 300.0°C

Cu10 WEED

*4

–200.0 to 300.0°C

Cu10 BAILEY

*4

–200.0 to 300.0°C

J263B

0.0 to 300.0K

±(0.05% of rdg + 0.3K)

±(0.1% of rdg + 1.5K)

0.1K

DI

Level

Vth = 2.4 V

Threshold level accuracy: ±0.1 V

Contact input

ON for 1 kΩ or less, OFF for 100 kΩ or more

*5

RTD

*7

(Measurement

current: 0.25 mA)

Pt500

*6

–200.0 to 600.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

0.1°C

JPt1000

*6

–200.0 to 600.0°C

Resistance

*7

20 Ω

(Measurement

current 1 mA)

0.000 to 20.000 Ω

±(0.05% of rdg + 7 digits)

±(0.1% of rdg + 25 digits)

0.001 Ω

200 Ω

(Measurement

current 1 mA)

0.00 to 200.00 Ω

±(0.05% of rdg + 3 digits)

±(0.1% of rdg + 15 digits)

0.01 Ω

2 kΩ

(Measurement

current 0.25 mA)

0.0 to 2000.0 Ω

±(0.05% of rdg + 3 digits)

±(0.1% of rdg + 10 digits)

0.1 Ω

*1 Pt50: JIS C 1604-1989, JIS C 1606-1989/Pt100: JIS C 1604-1997, IEC 751, DIN IEC 751/JPt100: JIS C 1604-1989, JIS C 1606-1989

*2 SAMA/DIN
*3 McGRAW EDISON COMPANY
*4 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/

Cu10 BAILEY: –20.0 to 250.0°C

*5 Measured using a measurement current of approximately 10 µA at 200-V range. Threshold level is approximately 0.1 V.

*6 The Pt500 resistance table is Pt100 × 5, and the resistance table for Pt1000 is Pt100 × 10.

*7 4-wire RTD, 4-wire resistance