beautypg.com

Index effects of operating conditions, External dimensions – Yokogawa PC-Based MX100 User Manual

Page 103

background image

4-15

IM MX100-01E

Specification

3

2

1

4

5

Index

Effects of Operating Conditions

The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:

At least 30 minutes after power-up

Effects of ambient temperature:

The effect received by changes in 10°C increments of the ambient
temperature is

±(0.05% of rdg + 0.05% of range) or less

However, ±(0.2% of range + 1 digit) for Cu10 Ω

Effects of power fluctuation:

Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V

Effects of magnetic field:

The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is

±(0.1% of rdg + 10 digits) or less

Effects from the signal source resistance:

Effects on DC voltage and thermocouple from a fluctuation of 1
kΩ in the signal source resistance

DC voltage: 1 V range or less

±10 µV or less

2 V range or higher

±0.15% of rdg or less

Thermocouple: ±10 µV or less

RTD:

Fluctuation due to 10 Ω change per cable (same
resistance for the three wires); 100 Ω system: ±0.1°
C or less, systems other than 100 Ω: ±1.0°C or less.
Fluctuation due to a difference in the resistance
between wires of 40 mΩ (maximum difference
among three wires) is approximately 0.1°C (for
Pt100)

Effect of position:

Horizontal with the feet at the bottom is the rule.

Effects of vibration:

The fluctuation that results by applying a sinusoidal vibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s

2

is ±(0.1% of rdg + 1 digit) or less

External Dimensions

Unit: mm

57

13

1

150.8

If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3

mm.