4-15
IM MX100-01E
Specification
3
2
1
4
5
Index
Effects of Operating Conditions
The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time:
At least 30 minutes after power-up
Effects of ambient temperature:
The effect received by changes in 10°C increments of the ambient
temperature is
±(0.05% of rdg + 0.05% of range) or less
However, ±(0.2% of range + 1 digit) for Cu10 Ω
Effects of power fluctuation:
Meets the accuracy specifications for AC power supply in the
range 90 to 132 V and 180 to 250 V
Effects of magnetic field:
The fluctuation in external magnetic fields for AC (50/60 Hz) 400
A/m is
±(0.1% of rdg + 10 digits) or less
Effects from the signal source resistance:
Effects on DC voltage and thermocouple from a fluctuation of 1
kΩ in the signal source resistance
DC voltage: 1 V range or less
±10 µV or less
2 V range or higher
±0.15% of rdg or less
Thermocouple: ±10 µV or less
RTD:
Fluctuation due to 10 Ω change per cable (same
resistance for the three wires); 100 Ω system: ±0.1°
C or less, systems other than 100 Ω: ±1.0°C or less.
Fluctuation due to a difference in the resistance
between wires of 40 mΩ (maximum difference
among three wires) is approximately 0.1°C (for
Pt100)
Effect of position:
Horizontal with the feet at the bottom is the rule.
Effects of vibration:
The fluctuation that results by applying a sinusoidal vibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is ±(0.1% of rdg + 1 digit) or less
External Dimensions
Unit: mm
57
13
1
150.8
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.
4.5 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10) Specifications