Index, Operation when the cf card is inserted, Normal operation – Yokogawa PC-Based MX100 User Manual
Page 128: When a power cut, When operating the access stop switch (cf-sw)
5-7
IM MX100-01E
Dual Save Functions (/DS Option)
3
2
1
4
5
Index
Operation When the CF Card Is Inserted
CF Write (Data Write Timing):
The timing at which data is written to a file (
CF File) on the CF card. Starting
with the second write operation (append), the operation is timed according to the
save interval. The write count is the count at which the file reaches 5 MB when
measuring at the shortest measurement interval. When the count is reached the file
closes, and a new file is created upon the next write operation. You can choose a
file creation mode of Rotary or FullStop mode (see “Data Storage” in section 4.2).
Info (File Information):
Files have numbers and start/stop conditions. You can access this information
using the PC software’s Viewer function.
• Normal Operation
Mode
Data
CF Write
CF File
Info
AlarmDO
Backup
Measure
Idle
Hold
Update
Update
5s-SW
5s-SW
5s-SW
FileNo.=0
FileNo.=1
Start
End
New file
Immediate save start Write count reached
Final save
Save interval
*
* Write count in this figure is an example only.
Count at which the file reches 5 MB at the shortest measurement interval.
(60 s)
Alarm Output (AlarmDO) is the same as when the CF card is not inserted.
• When a Power Cut
5s-SW
5s-SW
Mode
Data
CF Write
CF File
Backup
lost
PowerOff On
Backup
Measure
Measure
Auto restore
Start
Info
Abnormality New file
New file
End
FileNo.=0
FileNo.=1
FileNo.=2
(60 s)
• When Operating the Access Stop Switch (CF-SW)
5s-SW
5s-SW
Data
CF Write
CF File
Backup
FileNo.=0
lost
Info
CF -SW
5s-SW
Measure
Start
End
Measure
Idle
Mode
(60 s)
5.2 Explanation of Functions