Troubleshooting and Maintenance

3

2

1

4

5

Index

• When calibrating the RTD or resistance range of the 6-CH, Medium-Speed Four-Wire

RTD Resistance Input module

Short the input terminal

of CH3 (connect 0 )

When calibrating the RTD (1 mA)

20 mV range, connect 20 to the

input terminal of CH4.*

* When calibrating the RTD (1 mA) 60 mV range, connect 60

to the input terminal of CH5.

When calibrating the RTD (1 mA) 200 mV range, connect 200

to the input terminal of CH6.

When calibrating the RTD (1 mA) 600 mV range, connect 300

to the input terminal of CH4.

When calibrating the RTD (0.25 mA) 600 mV range, connect 2400

to the input terminal of CH5.

When calibrating the RTD (0.25 mA) 1 V range, connect 3000

to the input terminal of CH6.

Resistance standard

C

A

I

C

A

B

I

B

• When calibrating the range of the 4-CH, Medium-Speed Strain Module (-B12, -B35,

and -NDI)

The clamp terminal (-B12, -B35) and NDIS terminal (-NDI) are connected with the

4-gauge method (see below and next page). Resistors R1-R3 use 120 Ω, and a
resistance corresponding to the Zero or Full value is connected to R4.

To correctly calibrate the range, do so in the order Zero, then Full.

Calibration Range Res. R1, R2, and R3

Res. R4

Resistance value

accuracy

Zero

Full*

2000 µstrain

120.000 Ω

120.000 Ω

117.154 Ω

±0.005%, ±0.3ppm/°C

20000 µstrain

120.000 Ω

120.000 Ω

113.010 Ω

±0.005%, ±0.3ppm/°C

200000 µstrain

120.000 Ω

120.000 Ω

80.000 Ω

±0.005%, ±0.3ppm/°C

*

The Full calibration value of 2000 µstrain range is approximately to 12000 µstrain
The Full calibration value of 20000 µstrain range is approximately to 30000 µstrain
The Full calibration value of 200000 µstrain range is approximately 200000 µstrain

Setting switch (CH1 to CH4)

ON

OFF

No.1

No.2

No.3

No.4

No.5

For -B12, -B35

No.1

OFF

No.2

OFF

No.3

OFF

No.4

OFF

No.5

ON

Connect CH2

only

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

A(+)

B(L)

C(-V)

D(H)

R2

R4

R3

R1

3.3 Calibration