Module errors, Media-related errors, Communication errors – Yokogawa PC-Based MX100 User Manual
Page 74
3-2
IM MX100-01E
Module Errors
The left and right digits of the 7-segment LED display “E” and an error code,
respectively. The LED illuminates. In the case of module errors, the error number and the
corresponding module number are displayed alternately as shown in the figure below.
Error number
Module number
Display Probable Cause
Corrective Action
Ref. Page
E0
Range information error.
Servicing required.
-
E1
Calibration value error.
Check the module’s installation status, then recalibrate the
module.
If the error occurs even after recalibrating, servicing is required.
*
E2
Calibration reference voltage value is
not correct (during calibration).
Check whether the correct calibration reference voltage is being
applied or whether the channel to which the voltage is applied is
correct.
*
E3
Error in writing the calibration value.
Servicing required.
-
E4
The installed module cannot be used. Replace the module with one that can be used.
-
E5
Initial balancing Error
(Initial unbalance value adjustment)
Check the module’s installation status, then perform initial
balancing again.
If the error occurs even after recalibrating, servicing is required.
-
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
Media-Related Errors
The left and right digits of the 7-segment LED display “P” and an error code, respectively.
The LED blinks. The error code is saved to the log file (see “Log File Storage” under
“Saving Data to the CF Card” in section 1.2).
Display Probable Cause
Corrective Action
Ref. Page
P0
Manipulated the CF card while it was
being accessed.
Do not manipulate the CF card while it is being accessed.
2-27
P1
CF card error.
The CF card may not be formatted or may be damaged. Check
the CF card.
-
P2
The CF card was not formatted
correctly.
Insert the CF card into the CF card slot and format the card again,
or replace with another CF card.
*
P3
Not enough free space on the CF
card.
Delete unneeded data to increase the free space, or replace with
another CF card.
*
P4
The number of files saved on the CF
card exceeded 512.
Delete unneeded files to decrease the number of files, or replace
with another CF card.
*
P5
Overwrite error in the data storage
buffer.
The time needed to store the file must be reduced. Delete
unneeded data to increase the free space, or format the CF card.
If you are using a software application through the MX100/
DARWIN API, the timeout setting may not be appropriate. If so,
change the value to an appropriate value.
*
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
Communication Errors
The left and right digits of the 7-segment LED display “C” and an error code, respectively.
The LED blinks.
Display Probable Cause
Corrective Action
Ref. Page
C1
Multiple connection error.
Disconnect the current connection and reconnect. However, this
error may also appear when the Ethernet cable is removed and
inserted again. Turn OFF the power and reconnect.
*
*
See the MX100 Standard Software User’s Manual (IM MX180-01E).
3.1 Error Display on the 7-Segment LED and Corrective Actions