1.7 Functions of the 4-CH, Medium-Speed Strain
Input Module (-B12, -B35, and -NDI)
This module is equipped with four inputs and measures strain from strain gauges and
from strain gauge type sensors at a minimum measurement interval of 100 ms.
Input terminal
(clamp terminal)
Terminal cover
NDIS terminal
-NDI
-B12, -B35
Measurement Range
Input
Measurement Range Type
Rated Measurement Range
Strain
2000 µ Strain
–2000.0 to +2000.0 µ Strain
20000 µ Strain
–20000 to +20000 µ Strain
200000 µ Strain
–200000 to +200000 µ Strain
Measurement Interval, Integration Time, and Filter
Measurement Interval
Integration Time
Filter
Rejected Noise and Notes
100 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
200 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply
frequency and set 16.67 or 20 ms
500 ms
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
1 s
100 ms
Rectangular
10 Hz and its integer multiples
2, 5,10, 20, 30, 60 s
200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In such
cases, set the measurement interval to 200 ms or higher.
Measurement Synchronization
The module is equipped with a single A/D converter, and measurement is made
sequentially. Therefore, measurements of each channel are not synchronized.
Common Mode Voltage
For -B12 or -B35
The common mode voltage between channels in a single module when applied
continuously is 30 VACrms, or 250 VACrms between input and ground.
For -NDI
The common mode voltage between channels in a single module when applied
continuously is 30 VACrms, and 30 VACrms between input and ground (shell is non-
isolated).