Yokogawa Button Operated MV2000 User Manual
Page 96
3-16
IM MV1000-01E
• Number of set ponts > Mes val, True val
Press the
Input soft key and enter the value.
For
Mes val (measured value), set a value that is greater than the previous value.
Press the
Measure soft key to set the current measured value to Mes val. If you press
the
Measure soft key when you are setting the calibration function for more than one
channel, the measured value of the channel set to
Frst-CH is used as the Mes val for
all channels.
Range of Selectable Values for Mes val and True val
• On Channels Set to Lnear Scalng
–30000 to 30000 (decimal place is the same as that set for the scale)
• On All Other Channels
The measurable range
Example: If the range is 2 V, –2.0000 to 2.0000
Note
• If you change the Mode or Range settings, calibration correction (the
Number of set ponts
setting) is switched Off.
• You cannot set calibration correction for a channel that is set to Skip.
3.9 Calbratng Input Values (/CC1 Opton)