3.9 Calbratng Input Values (/CC1 opton)
This section explains how to calibrate input values to produce measured values.
For an explanation of this function, see section 1.2.
Dsplay
Press
MENU and then select > Menu tab > Meas channel > Calbraton correcton.
Settngs
True val
Number of set points
Calibration set points
Y5
Y4
Y3
Y2
Y1
MES val
X1 X2 X3 X4
X5
Condition
X1 < X2 < X3 < X4 < X5
• Frst-CH, Last-CH
Select the target channels. You can set consecutive channels whose range is set to
the same value as that of the first channel.
• Number of set ponts
Select the number of points (including the start and end points) that will be used to
divide segments, from within the range of 2 to 16.
To disable calibration correction, select Off.