13-16

IM MV1000-01E

Item

Specfcatons

Maximum input power

consumption

125 VA

Power consumption

Supply voltage

Minimum

Normal

Maximum

100 VAC

105 VA

105 VA

115 VA

240 VAC

105 VA

115 VA

125 VA

The operating conditions of the MV1000 and MV2000 are as follows:
Minimum: No USB connection, no optional terminal, back light saver ON, Auto save ON
Normal: No USB connection, no optional terminal, LCD brightness 2, Auto save ON
Maximum: No USB connection, with optional terminals, LCD brightness 8, Auto save ON

Dielectric strength

1500 VAC (50/60 Hz) for one minute between the AC adapter power supply line and earth

Miscellaneous

A/D converter integration time: Fixed to 20 ms (50 Hz) if set to Auto

Isolaton

Item

Specfcatons

Insulation resistance

Between the Ethernet, RS-422/485, and insulation terminals and earth: 20 MΩ or greater at 500

VDC

Dielectric strength

Between the power terminal and earth: 2300 VAC at 50/60 Hz for one minute
Between the contact output terminal and earth: 1600 VAC at 50/60 Hz for one minute
Between the signal input terminal and earth: 1500 VAC at 50/60 Hz for one minute
Between signal input terminals: 1000 VAC (50/60 Hz) for one minute (excluding the “b” input

terminal of MV1006, MV1012, MV1024, MV2010, MV2020,

MV2030, MV2040, MV2048)

Between the remote input terminal and earth: 1000 VDC for one minute
Between the pulse input terminal and earth: 1000 VDC for one minute

Ground

Grounding resistance: 100 Ω or less

Transportng and Storage Condtons

Item

Specfcatons

Ambient temperature

–25 to 60°C

Ambient humidity

5 to 95%RH (no condensation)

Vibration

10 to 60 Hz, 4.9 m/s

2

or less

Shock

392m/s

2

or less (in packaged condition)

Complant Standards

Item

Specfcatons

CSA

CSA22.2 No.61010-1, installation category II,

1

pollution degree 2,

2

and measurement category II

UL

UL61010-1 (CSA NRTL/C)

CE

EMC directive

EN61326 compliant (Emission: Class A, Immunity: Annex A)
EN61000-3-2 compliant
EN61000-3-3 compliant
EN55011

Low voltage directive EN61010-1, installation category II, measurement category II,

3

pollution degree 2

C-Tick

AS/NZS CISPR11 compliant, Class A Group 1

1 Installation category (overvoltage category) II: Describes a number which defines a transient overvoltage condition. Implies

the regulation for impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation
like a distribution board.

2 Pollution degree 2: Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface

resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.

3 Measurement category II: Applies to measuring circuits connected to low voltage installation, and electrical instruments

supplied with power from fixed equipment such as electric switchboards.

13.6 General Specfcatons

This manual is related to the following products: