Difference Computation

Low-cut

Square Root Computation

Calibration Correction (/CC1 Option)

Input

value

Measured

value

Measured

value

Input

value

Input

value

Input

value

Input

value

Measured

value

Measured

value

Measured value on the reference channel

The measured value of the channel is set to the

difference with respect to the measured value of the

reference channel.

Converts the unit to obtain the measured value.

Takes the square root of the input value and converts

the unit to obtain the measured value.

Corrects the input value with the characteristics

specified by segments to obtain the measured value.

Number of break

points: Up to 16

Pulse Input (/PM1 Option)

Linear Scaling

10 V

0 V

300 °c

–100 °c

a

b

x

a

b

B

B

X

A

A

X = (B – A)

+ A

x – a

b – a

The contact changes from open to closed.

The signal level at the input terminal changes from high

to low.

5 ms or more

Contact

or

Open

collector

Pulses of 100 Hz or less

can be counted

Counts pulses.

Input to the MV

Count on the MV

Measured

value

Output

value

For square root computation, measured values below

the specified value are cut.

For 1-5 V input, values below 0 % are cut.

Result of square root

computation

Low-cut value

MV

For configuration instructions, see section 3.3.

For configuration instructions, see section 3.3.

For configuration instructions, see section 3.9.

For configuration instructions, see section 3.3.

For configuration instructions, see section 3.3.

For configuration instructions, see section 3.10.

( )

Note

You can use difference computation even if the reference channel and difference computation
channel’s input types and ranges are different. Difference computation takes place using the
difference computation channel’s decimal place and unit.

Example 1: If the input value of the difference computation channel is 10.00 and the measured

value of the reference channel is 100.0, the result is 10.00–100.0=–90.00.

Example 2: If the input value of the difference computation channel is 10.00 V and the

measured value of the reference channel is 5.00 mV, the computed result is 10.00
V – 5.00 mV = 5.00 V.

1.2 Input