Computaton Functon (ncludng the report functon) (/M1)
Item
Specfcatons
Number of computation channels MV1004, MV1008, MV2008: 12 channels (101 to 112)
MV1006, MV1012, MV1024: 24 channels (101 to 124)
MV2010, MV2020, MV2030, MV2040, MV2048: 60 channels (101 to 160)
Operation
General arithmetic operations: Basic arithmetic, square root, absolute, common logarithm,
natural logarithm, exponential, and power
Relational operations: <, ≤, >, ≥, =, ≠
Logic operations: AND, OR, NOT, XOR
Statistical operations: TLOG, CLOG
Special operations: PRE, HOLD, RESET, CARRY
Conditional operation: [a?b:c]
Computation accuracy
Double-precision floating point
Usable data
Channel data
Measurement, computation, and external input channels (/MC1)
Constants
60 constants
Communication input data
MV1000: 24, MV2000: 60
Remote control input status
0/1 (/R1)
Pulse input
Counts the number of pulses (/PM1)
Status input
Internal switch, alarm output relay (/A[ ]), and flags
Rolling average:
Performs moving average on the computed results.
Measure range
–9999999 to 99999999
Decimal place: Zero to four fractional digits
Alarms
High limit, low limit, delay high limit, and delay low limit
Hysteresis: High and low limit alarm: 0.0% to 5.0% of the span.
Display
Same as the measurement channels
Data saving
Same as the measurement channels
Report function
Number of report channels: same as the number of computation channels
Operations: Average, maximum, minimum, sum, instantaneous value
Report types: Hourly, daily, hourly + daily, daily + weekly, daily + monthly
Cu10, Cu25 RTD Input/Three-Wre Isolated RTD Input (/N1)
Item
Specfcatons
Function
In addition to the standard input, the MV can also receive Cu10 and Cu25 input.
On the MV1006, MV1012, MV1024, MV2010, MV2020, MV2030, MV2040, and MV2048,
all the RTD input terminals (A, B, and b) are isolated on each channel.
Measurement/display accuracy
Under standard operating conditions
Input Type
Measurement
Range
Accuracy
Guaranteed
Range
Measurement Accuracy
Max.
Resolution
A/D integration time:
16.7 ms or more
A/D integration time:
1.67 ms
Cu10 (GE)
Cu10 (L&N)
Cu10 (WEED)
Cu10 (BAILEY)
Cu10: α = 0.00392 at 20°C
Cu10: α = 0.00393 at 20°C
Cu25: α = 0.00425 at 0°C
–70 to 170°C
–75 to 150°C
–200 to 260°C
–200 to 300°C
–328 to 572°F
–200 to 300°C
* Measuring current i = 1 mA
(0.4% of rdg + 1.0°C)
(0.3% of rdg + 0.8°C)
(0.8% of rdg + 5.0°C)
(0.5% of rdg + 2.0°C)
0.1°C
Input source resistance
1 Ω or less per wire (The resistance of all three wires must be equal.)
Ambient temperature influence (applies when the A/D integration time is 16.67 ms or greater)
±(0.2% of range + 2 digits) or less
Signal source resistance
With a variation of 1 Ω per wire (same resistance for all three wires): ±(0.1% of rdg +
1 digit) or less
With a maximum difference of 40 mΩ between wires: Approx. 1°C
13.5 Optons