12.4 Calbratng the MV

It is recommended that the MV be calibrated once a year to assure its measurement
accuracy.
Calibration service is also provided by YOKOGAWA dealers.
For details, contact your nearest YOKOGAWA dealer.

Requred Instruments

Calibration instruments with the following resolution are required for calibrating the MV.

Recommended Instruments

• DC voltage standard:

9100 by FLUKE or equivalent

Main specifications

Output accuracy: ±(0.005% + 1 µV)

• Decade resistance box:

Yokogawa Meters & Instruments Model 2793-01
or equivalent

Main specifications

Accuracy of output range 0.1 to 500 Ω: ±(0.01%
+2 mΩ)

Resolution: 0.001 Ω

• 0°C standard temperature device: ZC-114/ZA-10 by Coper Electronics or equivalent

Main specifications

Standard temperature stability accuracy: ±0.05°C

For information on purchasing the calibration instruments, contact your nearest
YOKOGAWA dealer.

Calbraton Procedure

1.

Wire the MV and the calibration instrument as shown in the following figure, and

adequately warm up the instruments (the warm-up time of the MV is at least 30
minutes).

2.

Check that the operating environment such as ambient temperature and humidity

is within the standard operating conditions (see section 13.6).

3.

Apply appropriate input signals corresponding to 0, 50, and 100% of the input

range and calculate the errors from the readings.

If the error does not fall within the accuracy range of the specifications, contact
your nearest YOKOGAWA dealer.

Note

For thermocouple inputs, you must measure the temperature of the input terminal and apply a
voltage taking into account the reference junction temperature.

DC Voltage Measurement (Example for the MV1012)

DC voltage standard

Input terminal

Screw terminal

Clamp terminal

+

+

CH1

CH2

CH3

CH4

CH5

CH6

CH7

CH8

CH9

CH11

CH10

CH12

+/A

/b

–/B

CH5 CH3 CH1

CH6 CH4 CH2

CH9 CH7

CH10

CH11

CH12

CH8

/b

+/A

–/B

This manual is related to the following products: