7.1 Confgurng the Event Acton Functon

This section explains how to configure the MV to execute a specified action when a
given event occurs. This function is called event action. This section also explains how to
configure the remote control function (/R1 option) and the USER key.

For an explanation of these functions, see section 1.7.
For event action configuration examples, see appendix 2.

Dsplay

Press

MENU and then select Menu tab > Tmer, Event acton > Event acton.

Settngs

• Logc box number

You can set up to 40 event actions.

• Event

An event is a condition for executing an action.

Settng

Descrpton

None

Remote

Select a remote control input terminal number.

Relay

Select an alarm output relay number.

Switch

Select an internal switch number.

Timer

Select a timer number.

Match T

A match time timer.
Select a match timer number.

Alarm

UserKey

• Acton

An action that is executed when an event occurs.

Settng

Descrpton

MV1000

MV2000

Memory

Memory

Memory start/stop

Start

Start

Memory start

Stop

Stop

Memory stop

Trigger

Trigger

An event trigger.
You can specify this when the MV is configured to record

event data.

AlarmACK

AlarmACK

You cannot specify this when the event is set to Relay,

Switch, or Alarm.

Math

Math

Computation start/stop. You can specify this on models with

the /M1 or /PM1 option.

MathStart

MathStart

You can specify this on models with the /M1 or /PM1 option.

MathStop

MathStop

You can specify this on models with the /M1 or /PM1 option.

MathReset

MathReset

You can specify this on models with the /M1 or /PM1 option.