beautypg.com

5 saving measured data manually, 5 saving measured data manually -13, 5 savng measured data manually – Yokogawa Button Operated MV2000 User Manual

Page 113

background image

4-13

IM MV1000-01E

Measurement and Record

ng

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

4.5 Savng Measured Data Manually

This section explains how to save the current values of all channels (except for channels
set to Skip or Off) through key operations.

For an explanation of this function, see section 1.5.

Procedure

1.

In Operation Mode, press

FUNC.

The function menu appears.

2.

Press the

Manual sample soft key.

Manual sampling begins.

Explanaton

• Manually Sampled Data n Internal Memory

You can check the number of manually sampled data files in internal memory on the

Memory Summary display (see section 1.4 for details).

• Savng to a CF Card

• If

Auto save is set to On, manually sampled data will be automatically saved to a

CF card when manual sampling takes place.

• If

Auto save is set to Off, use a manual save operation (see section 4.4 for details)

to save manually sampled data to a CF card or to USB flash memory.

• No matter what

Auto save is set to, you can always save manually sampled data to

a CF card or to USB flash memory using a manual save operation (see section 5.9
for details).

Settng Whch Channels to Save Manually (only on the MV2000)

On models of the MV2000 equipped with external input channels (/MC1 option), you can
set which channels’ (120 channels or less) current values to save.

Dsplay

Available on models equipped with external input channels (/MC1 option).
Press MENU and then select > Menu tab > Data save > Manual sample.

This manual is related to the following products: