1.2 Input

Measurement Channels

• Number of Channels vs. Scan Interval

The MV acquires data by sampling measurement channel input signals at the set scan

interval. The following is a list of the scan intervals available with different numbers of

channels.

Model

Number of

Measurement

Channels

Scan Interval

Normal Mode

Fast Samplng Mode

MV1004

4

125 ms, 250 ms

25 ms

MV1008

8

MV1006

6

1 s, 2 s, 5 s

2 s, 5 s

125 ms

MV1012

12

MV1024

24

MV2008

8

125 ms, 250 ms

25 ms

MV2010

10

1 s, 2 s, 5 s

2 s, 5 s

125 ms

MV2020

20

MV2030

30

MV2040

40

MV2048

48

A/D converter integration time

60 Hz/50 Hz

100 ms

600 Hz (fixed)

For configuration instructions, see section 3.1.

• A/D Converter Integraton Tme

The MV uses an A/D converter to convert sampled analog signals to digital signals.

You can effectively eliminate the influence of power supply noise by setting the A/D

converter’s integration time to the same length as or to an integral multiple of the

power source period.

• Because 100 ms is an integral multiple of both 16.7 ms and 20 ms, you can effectively

eliminate the influence of power supply noise from both 50- and 60-Hz frequencies by
setting the integration time to 100 ms.

• Power supply noise elimination is less effective in fast sampling mode than it is in normal

mode. When measuring in an environment susceptible to power supply noise, we
recommend that you measure in normal mode.

For configuration instructions, see section 3.1.

Input Types and Computaton Functons

You can measure the following types of input.

Input Type

Descrpton

DC voltage

You can measure DC voltages in the range of ±20 mV to ±50 V.

DC current

You can measure a DC current signal by converting it to a voltage signal using a shunt

resistor attached to the input terminal.

1

The converted signal can be measured within the DC voltage range (see above).

Thermocouple You can measure temperatures using these thermocouple types: R, S, B, K, E, J, T, N,

W, L, U, and WRe3-25. It is also possible to measure using other thermocouples, such as

PR40-20 and PLATINEL.

2

RTD

You can measure temperatures using RTD types Pt100 and JPt100. It is also possible to

measure using other RTD types such as Cu10 and Cu25,

3

and Pt50 and Ni100.

2

ON/OFF input You can display contact input or voltage input signals correlated to 0% or 100% of the

display range.
Contact input: A closed contact is ON (1). An open contact is OFF (0).
Voltage input: Less than 2.4 V is OFF (0). 2.4 V or more is ON (1).

Pulse input

4

You can count pulses.

1 Shunt resistor sold separately. For example, you can use a 250-Ω shunt resistor to convert a 4- to 20-mA

signal to a 1- to 5-V signal.

2 /N3 option
3 /N1 option
4 /PM1 option