2 setting timers, 2 setting timers -3, 2 settng tmers – Yokogawa Button Operated MV2000 User Manual
Page 194
7-3
IM MV1000-01E
Event
Act
on
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
7.2 Settng Tmers
You can set timers on the events of the event action function. Timers are also used in
TLOG computation.
For an explanation of the event action function, see section 7.1.
Dsplay
Press
MENU and then select Menu tab > Tmer, Event acton > Tmer.
• When relative time is selected
• When absolute time is selected
Settngs
You cannot change these settings during memory sampling or computation.
• Tmer No.
You can set four timers (1 to 4).
When Usng a Relatve Tmer
• Mode
Select
Relatve.
• Interval
Set the interval in the range of 00:01 (1 minute) to 24:00 (24 hours).
• Reset at Math Start
On: Resets the timer when computation is started. The resetting of the timer is not
considered a timeout. Even if the timer is used as an event, the action is not executed.
When Usng an Absolute Tmer
• Mode
Select Absolute.
• Interval
Set the interval from within the range of available settings (1min to 24h).
• Ref. tme
Set the time in the range of hour 0 to hour 23.