6 general specifications, 6 general specifications -14, 6 general specfcatons – Yokogawa Button Operated MV2000 User Manual
Page 282: Constructon, Normal operatng condtons, Input termnal (nput termnal suffx code -1), Input termnal (nput termnal suffx code -2)
13-14
IM MV1000-01E
13.6 General Specfcatons
Constructon
Item
Specfcatons
Material
Case: Aluminum plate (MV2000), metal plate (MV1000)
Bezel: Polycarbonate
Color
Case: Smoke blue (Munsell 4.1PB6.0/4.5 or equivalent)
Bezel: Light gray (Munsell 5.2PB8.2/1.0 or equivalent)
External dimensions
MV1000: 189(W) × 177(H) × 253(D) mm, 189(W) × 177(H) × 259(D) mm
*1
, 189(W) × 186(H) ×
259(D) mm
*2
MV2000: 307(W) × 273(H) × 254(D) mm , 307(W) × 273(H) × 260(D) mm
*1
*1 With clamp terminals (input terminal suffix code is -1) or pulse inputs (/PM1 option)
*2 Rechargeable battery model (power supply suffix code is -3)
Weight
MV1004, MV1006, MV1012: Approx. 3.3 kg, Approx. 4.5 kg
*1
, MV1008, MV1024: Approx. 3.5 kg,
Approx. 4.7 kg
*1
MV2008, MV2020: Approx. 5.3 kg, MV2010: Approx. 5.2 kg, MV2030: Approx. 5.5
kg, MV2040, MV2048: Approx. 5.6 kg excluding options
*1 Rechargeable battery model (power supply suffix code is -3)
Normal Operatng Condtons
Item
Specfcatons
Supply voltage
90 to 132, 180 to 250 VAC
Power supply frequency
50 Hz ± 2%, 60 Hz ± 2%
Ambient temperature
0 to 40°C, 10 to 40°C (Rechargeable battery model (power supply suffix code is -3))
Ambient humidity
20 to 80%RH (at 5 to 40°C)
Vibration
10 to 60 Hz, 0.2 m/s
2
Shock
Not allowed
Magnetic field
400 A/m or less (DC and 50/60 Hz)
External noise
Normal mode (50/60 Hz)
DC voltage
The peak value including the signal must be less than 1.2 times the measuring range.
Thermocouple
The peak value including the signal must be less than 1.2 times the measuring thermal
electromotive force.
RTD
50 mV or less
Common mode noise
250 VACrms or less for all ranges (50/60 Hz)
Maximum noise voltage between channels
250 VACrms (50/60 Hz) or less
Installation position
Horizontal
You can tilt the MV by bringing the front legs out.
Warm-up time
At least 30 minutes after power-on
Installation location
Indoors
Operating altitude
2000 m or less
Input Termnal (nput termnal suffx code -1)
Item
Specfcatons
Type
Clamp terminal
Attachable and detachable for each channel
Recommended wire size: 0.08 mm
2
to 1.5 mm
2
(AWG28 to 16)
Input Termnal (nput termnal suffx code -2)
Item
Specfcatons
Type
M4 screw