App index – Yokogawa Button Operated MV2000 User Manual
Page 254
12-9
IM MV1000-01E
Troubleshoot
ng and Ma
ntenance
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Code
Message
285
FTP data connection error.
Further details are provided by the character string that appears after error code 285.
Character Strng and Detals
SOCKET
Failed to create a socket for the transfer connection.
2
BIND
Failed the transfer connection command.
2
CONNECT
Failed the transfer connection.
2
LISTEN
Failed the transfer connection reception.
2
ACCEPT
Failed to accept the transfer connection.
2
SOCKNAME
Internal processing error.
1
RECV
Failed to receive data over the transfer connection.
2
SEND
Failed to send data over the transfer connection.
2
286
FTP file transfer error.
290
SNTP access failure.
Further details are provided by the character string that appears after error code 290.
Character Strng and Detals
DORMANT
Internal processing error.
1
LINK
Data link is disconnected.
Check the cable connection.
291
SNTP server does not respond.
Further details are provided by the character string that appears after error code 291.
Character Strng and Detals
TIMEOUT
Check that the server is running.
2
292
Incorrect SNTP server setting.
Further details are provided by the character string that appears after error code 292.
Character Strng and Detals
HOSTNAME
Failed the DNS lookup (search the IP address corresponding to the host name).
Check the DNS setting and the SNTP server name.
TCPIP
Internal processing error.
1
293
Invalid SNTP server reply.
Further details are provided by the character string that appears after error code 293.
Character Strng and Detals
SEND
A correct IP address has not been assigned to the MV.
Check the IP address.
BROKEN
There is a problem with the SNTP server.
If this error occurs even after executing SNTP manually several times, check the SNTP
server.
12.1 Lst of Messages