beautypg.com

3 alarms, 3 alarms -6, Alarm types – Yokogawa Button Operated MV2000 User Manual

Page 15

background image

1-6

IM MV1000-01E

1.3 Alarms

This function generates an alarm when the measured data meets a set condition. You

can set a maximum of four different alarms on each channel.

Alarm Types

You can use the types of alarms listed below. The letters contained in parentheses are

used as abbreviations for the alarms.

For alarm configuration instructions, see section 3.7.

High Limit Alarm (H)

Alarm value

Alarm output ON

Alarm output ON

Alarm output ON

Alarm output ON

Alarm value

Hysteresis

Measured value

Measured value Hysteresis

Low Limit Alarm (L)

Alarm output ON

Alarm output ON

Delay time

Delay time

Alarm value

Alarm value

Measured value

Measured value

Delay High Limit Alarm (T)

Delay Low Limit Alarm (t)

High Limit on Rate-of-Change Alarm (R)

Low Limit on Rate-of-Change Alarm (r)

Alarm value

Alarm value

Hysteresis

Hysteresis

Difference in the measured values of two channels

Difference in the measured values of two channels

Difference High Limit Alarm (h)

Difference Low Limit Alarm (l)

Change in the

measured value

Change in the measured value

Measured value

Measured value

Time

Time

Interval (t2-t1)

Interval (t2-t1)

Amount of change in

the setting
|

|

T2

T1

t1

t2

T2

T1

t1

t2

T2-T1

Amount of change in

the setting

|

T2-T1

|

• Hysteress

You can set a difference between the value that activates an alarm and the value that

deactivates it.

For configuration instructions, see section 3.5.

• Delay Hgh Lmt Alarm and Delay Low Lmt Alarm

If the measured value remains above or below the set alarm value for the set period of

time (the delay time), an alarm is activated.

• Hgh Lmt on Rate-of-Change Alarm and Low Lmt on Rate-of-Change

Alarm

The MV checks the rate-of-change of the measured values over a set period of time (the

interval) and activates an alarm if the rate-of-change in the rising or falling direction is

greater than or equal to the set value.

The alarm value of the rate-of-change alarm is set using an absolute value. The

interval is derived from the sampling count using the following equation:

Interval = scan nterval × samplng count

For instructions on how to set the interval, see section 3.5.

• Dfference Hgh Lmt Alarm and Dfference Low Lmt Alarm

The MV activates an alarm when the difference between the values of two channels

goes above or below the set value. You can use this alarm on measurement channels

set to difference computation.

This manual is related to the following products: