Yokogawa Button Operated MV2000 User Manual
Page 216
9-8
IM MV1000-01E
9.6 Transmttng the MV Status va Relay Contact
(/F1 opton)
This section explains how to configure the MV to transmit a relay contact signal when an
error occurs in the MV CPU. It also explains how to configure the MV to transmit a signal
to another relay when a specific condition occurs.
For an explanation of these functions, see section 1.10.
Dsplay
Press
MENU and then select Menu tab > Basc settng mode > Menu tab > Status
relay.
Settngs
• Memory/Meda status
On: Transmits the internal memory and CF card statuses to a relay.
• Measurement error
On: Transmits a relay signal when a measurement error occurs.
• Communcaton error
On: Transmits a relay signal when a communication error occurs.
• Memory stop
On: Outputs a relay signal when memory sampling is stopped.
Procedure
• FAIL Output
No settings or operations are required. The MV transmits a relay contact signal if it
detects a CPU error. The MV also transmits a relay contact signal when the MV is
turned OFF.
• Status relay
The MV transmits a relay contact signal when a specified condition occurs.