9 setting display item, Setting display item -9, Maintenance – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 97: Display setup
<7. Startup>
7-9
IM 11M12A01-02E
7.9
Setting Display Item
This section briefly describes the display item settings shown in Figure 7.11, “Basic Panel
Display.”
F7.11E.ai
Tag:
21.0
%O
2
17.43mA -Output1
17.43mA -Output2
Primary value
Secondary value
Tertiary value
Figure 7.11 Basic Panel Display
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” in the Execution/Setup display.
(2) Select the “Display setup” from the Maintenance panel display (Figure 7.12).
The Display setup display (Figure 7.13) then appears.
(3) In the above Display setup display, select the “Display item”. The Display item display
(Figure 7.14) then appears. From this display, select the “Primary value” and press the
[Enter] key to display the “Display item” selection display (Figure 7.15).
(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, Display Items, enabling the selection of display items in individual display
areas.
Enter
Display setup
Calibration setup
Blow back setup
mA-output loop check
Contact check
Maintenance
F7.12.ai
Enter
Display item
Trend graph
Auto return time: 0min
Language: English
Display setup
F7.13.ai
Figure 7.12 Maintenance Panel Display
Figure 7.13 Display Setup
8th Edition : Jan.13,2012-00