background image

<12. Troubleshooting>

12-1

IM 11M12A01-02E

8th Edition : Jan.13,2012-00

12. Troubleshooting

This chapter describes errors and alarms detected by the self-diagnostic function of the

converter. It also explains inspections and remedies when other problems occur.

12.1 Displays and Remedies When Errors Occur

12.1.1 Error Types

An error occurs when an abnormality is detected in the detector or the converter, e.g., in the cell

(sensor), detector heater, or internal circuits of the the converter..

If an error occurs, the converter performs the following:
(1) Stops the supply of power to the heater in the detector to insure system safety.
(2) Causes an error indication in the display to start blinking to notify of an error generation

(Figure 12.1).

(3) Sends an output contact if the error is set up for Output contact setup for that contact

(refer to Section 8.4, Output Contact Setup).

(4) Changes the analog output status to the one set in Output hold setting.

(refer to Section 8.2, Output Hold Setting).

When the display shown in Figure 12.1 appears, pressing the error indication brings up a

description of the error (Figure 12.2). The content of errors that are displayed include those

shown in Table 12.1.

Tag:

21.0

%O

2

17.42mA -Output1

17.42mA -Output2

Hold

F12.1E.ai

Tag:

Error

2 :Heater temp.

Alarm 1 1 : T.C. voltage

Hold

F12.2E.ai

Figure 12.1

Figure 12.2

Table 12.1 Types of Errors and Reasons for Occurrence

Error Code

Error Type

Occurrence Conditions

Error1

Cell voltage failure

The cell (sensor) voltage signal input to the converter falls below -50 mV.

Error2

Heater temperature failure The heater temperature does not rise during failure warm-up, or it falls

below 730 ºC or exceeds 780 ºC after warm-up is completed.

Or this occurs if the TC+, TC- thermocouple terminals are wired to

converter with reverse (wrong) polarity.

Error3

A/D converter failure

The A/D converter fails in the internal electrical circuit in the converter.

Error4

Memory failure

Data properly are not written into memory in the internal electrical circuit

in the converter.