11 checking contact i/o, 1 checking contact output, Checking contact i/o -12 7.11.1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 100: Checking contact output -12, Output contacts

<7. Startup>
7-12
IM 11M12A01-02E
7.11 Checking Contact I/O
Conduct the contact input and output checking as well as operational checking of the solenoid
valves for automatic calibration.
7.11.1 Checking Contact Output
To check the contact output, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Select “Maintenance” in that display.
(2) Select “Contact check” then “Output contacts” in the Maintenance panel display to display
the Output contacts display (see Figure 7.15.2).
(3) In this display, select the desired output contact for checking. The display, which enables
the closing and opening of contacts, then appears. Use the display to conduct continuity
checking.
Output contact 1 : Open
Output contact 2 : Open
Output contact 3 : Open
Output contact 4 : Open
Enter
Output contacts
F7.15.2E.ai
Figure 7.15.2 Output Contact Check Panel Display
NOTE
• If you conduct an open-close check for contact output 4, Error 1 or Error 2 will occur.
This is because the built-in heater power of the detector, which is connected to contact
output 4, is turned off during the above check. So, if the above error occurs, reset the
equipment or turn the power off and then back on to restart (refer to Section 10.4, Reset,
later in this manual).
8th Edition : Jan.13,2012-00