Specifications, 1 general specifications, 1 standard specifications – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 19: Specifications -1, General specifications -1 2.1.1, Standard specifications -1

<2. Specifications>
2-1
IM 11M12A01-02E
8th Edition : Jan.13,2012-00
2.
Specifications
This chapter describes the specifications for the following:
ZR22G
General use separate type detector (See Section 2.2.1)
ZO21R
Probe protector (See Section 2.2.2)
ZR22G (0.15 m)
High temperature separate type detector (See Section 2.3.1)
ZO21P
Adapter for High temperature probe (See Section 2.3.2)
ZR402G
Separate type converter (See Section 2.4)
ZA8F
Flow setting unit (See Section 2.5.1)
ZR40H
Automatic calibration unit (See Section 2.5.2)
ZO21S
Standard gas unit (See Section 2.6)
2.1
General Specifications
2.1.1
Standard Specifications
Measured Object:
Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases, may not be applicable corrosive gas such
as ammonia is present-check with YOKOGAWA)
The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.
Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O
2
Output Signal:
4 to 20 mA DC (maximum load resistance 550 Ω)
Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
2
(in 1 vol% O
2
), or partial range
Digital Communication (HART): 250 to 550 Ω, depending on number of field devices connected
to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range:
0 to 100 vol% O
2
Warm-up Time:
Approx. 20 min.
Repeatability:
(Excluding the case where the reference gas is by natural convection)
±0.5% Maximum value of set range. (range from 0 to 5 vol% O
2
or more
and less than 0 to 25 vol% O
2
range)
±1% Maximum value of set range.
(range from 0 to 25 vol% O
2
or more and up to 0 to 100 vol% O
2
range)
Linearity:
(Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection)
(Use oxygen of known concentration (within the measuring range) as the
zero and span calibration gases.)
±1% Maximum value of set range ;
0 to 5 vol% O
2
or more and less than 0 to 25 vol% O
2
range
(Sample gas pressure: within ±4.9 kPa)
±3% Maximum value of set range ;
0 to 25 vol% O
2
or more and less than 0 to 50 vol% O
2
range