3 default values, Default values -12 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 118
<8. Detailed Data Setting>
8-12
IM 11M12A01-02E
8.4.3
Default Values
When the analyzer is delivered, or if data are initialized, alarm and other setting defaults are as
shown in Table 8.6.
Table 8.6 Output Contact Default Settings
Item to be selected
Output
contact 1
Output
contact 2
Output
contact 3
Output
contact 4
Alarm
settings
High-high limit alarm
�
High limit alarm
ON
�
Low limit alarm
ON
�
Low-low limit alarm
�
Calibration coefficient alarm
�
Startup power stabilization
timeout alarm
�
Error
ON
Other
settings
Warm-up
ON
�
Output range change
�
Calibration
ON
�
Maintenance
ON
�
Blow back
�
High limit temperature alarm
�
Calibration gas press. low
(default)
(default)
(default)
�
Process upset
�
Operating contact status
Open
Closed
Closed
Closed
(fixed)
Note: Blank boxes in the above table indicate that the default is “disabled“.
8th Edition : Jan.13,2012-00