<9. Calibration>

9-7

IM 11M12A01-02E

9.2.1.5 Setting Calibration Time

• When the calibration mode is in manual:
First set the “Hold time” (output stabilization time). This indicates the time required from the

end of calibration to entering a measurement again. This time, after calibration, the sample

gas enters the sensor to set the time until the output returns to normal. The output remains

held after completing the calibration operation until the hold (output stabilization) time elapses.

The calibration time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. For

more details, consult Section 8.2, Output Hold Setting. When the calibration mode is in semi-

automatic, set the hold (output stabilization) time and calibration time. The calibration time is the

time required from starting the flow of the calibration gas to reading out the measured value. The

set calibration time is effective in conducting both zero and span calibrations. The calibration-

time set ranges from 00 minutes, 00 seconds to 60 minutes, 59 seconds. Figure 9.6 shows the

relationship between the calibration time and hold (output stabilization) time.

Analog output remains held

(when output remains held)

Calibration start

(contact or switch input)

Span calibration

(span gas valve open)

Zero calibration

(zero gas valve open)

Analog output status

Calibration time

Calibration time

Hold (output

stabilization) time

F9.6E.ai

Figure9.6 Calibration and Hold Time Settings

• When the calibration mode is in automatic:
In addition to the above Hold (output stabilization) time and Calibration time, set the Interval, Start

date, and Start time.
Interval means the calibration intervals ranging from 000 days, 00 hours to 255 days, 23 hours.
Set the first calibration day and the start-calibration time to the “Start date” and “Start time”

respectively. For example, to start the first calibration at 1:30 p.m. on March 25, 2001, enter

25/03/01 to the start date and 13 hours, 30 minutes to the start time, following the steps below:
(1) Select the “Calibration timing” display. A panel display as shown in Figure 9.7 appears.
(2) Select each item for the calibration to display the numeric-data entry display. Enter the

desired numeric values for the calibration.

Enter

Hold time: 1 0 min 0 0 s

Cal time: 1 0 min 0 0 s

lnterval: 3 0 d 0 0 h

Start date: 0 1 / 0 1 / 0 0

Start time: 0 0 : 0 0

Calibration timing

F9.7E.ai

Figure 9.7 Calibration Timing Display

8th Edition : Jan.13,2012-00