2 setting procedure, 3 default values, Setting procedure -14 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 120: Default values -14, Input contacts
<8. Detailed Data Setting>
8-14
IM 11M12A01-02E
8.5.2
Setting Procedure
This setting example shows how to set “When contact input 1 Opens, Start Semi_Auto
Calibration”.
1) From the Basic panel display touch [Setup] key, and the [Execution/Setup] display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Contact Setup” then “Input contacts”.
4) Select “Input 1” . Function Selection window (Fig 8.11) appears.
5) Select “Calibration start”.
6) For Input contact 1 select (operate if) “Closed”. (You can select “Open” or “Closed”).
7) Change Closed to Open.
Input1: No function
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
Input contacts
F8.10E.ai
Input1: No function
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
Input contacts
No function
Cal. gas press. low
Range change
Calibration start
Process up set
Blow back start
F8.11.ai
Figure 8.10 Input Contacts Display
Figure 8.11 Input Contacts Display
8.5.3
Default Values
All contact inputs are set to “No function”(disabled) prior factory shipment or after data
initialization..
8th Edition : Jan.13,2012-00