2 wiring for detector output, 1 cable specifications, Wiring for detector output -5 5.2.1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 77: Cable specifications -5, Caution
<5. Wiring>
5-5
IM 11M12A01-02E
8th Edition : Jan.13,2012-00
5.2
Wiring for Detector Output
This wiring enables the converter to receive cell output from the detector, output from a
thermocouple and a reference junction compensation signal. Install wires that allow for 10 Ω of
loop resistance or less. Keep detector wiring away from power wiring.
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
TC(-)
CJ(+)
CJ(+)
CJ(-)
CJ(-)
FG
Shielded cables
Shielded cables
ZR22G
Detector
ZR402G
Converter
CELL(+)
CELL(+)
CELL(-)
CELL(-)
TC(+)
TC(+)
TC(-)
TC(-)
CJ(+)
CJ(+)
CJ(-)
CJ(-)
FG
(2) Ambient temperature of the detector: exceeding 80 °C
Heat-resistant wiring
Terminal box
(1) Ambient temperature of the detector: 80 °C or less
F5-5E.ai
ZR22G
Detector
ZR402G
Converter
Figure 5.4 Wiring for detector output
CAUTION
If shielded cables cannot be used between the detector and the terminal box, for example,
when heat-resistant wiring is used, locate the detector and the terminal box as close together as
possible.
5.2.1
Cable Specifications
Basically, PVC insulated PVC sheathed cable (6-core) is used for this wiring. When the ambient
temperature of the detector exceeds 80 °C, install a terminal box, and connect with the detector
using six-piece 600 V silicon rubber insulated glass braided wires.