<2. Specifications>
2-7
IM 11M12A01-02E
8th Edition : Jan.13,2012-00
2.2.2
ZO21R Probe Protector
Used when sample gas flow velocity is approx. 10 m/sec or more and dust particles wears the
detector in cases such as pulverized coal boiler of fluidized bed furnace (or burner) to protect
the detector from wearing by dust particles. When probe insertion length is 2.5 m or more and
horizontal installation, specify the ZO21R-L-200-
□
*B to reinforce the probe.
Insertion Length:
1.05 m, 1.55 m, 2.05 m.
Flange:
JIS 5K 65A FF equivalent. ANSI Class 150 4 FF (without serration)
equivalent or DIN PN10 DN50A equivalent. However, flange thickness is
different.
Material:
SUS316 (JIS), SUS304 (JIS) (Flange)
Weight:
1.05 m; Approx. 6/10/8.5 kg (JIS/ANSI/DIN),
1.55 m; Approx. 9/13/11.5 kg (JIS/ANSI/DIN),
2.05 m; Approx. 12/16/14.5 kg (JIS/ANSI/DIN)
Installation:
Bolts, nuts, and washers are provided for detector, probe adapter and
process-side flange.
Model and Codes
Model
Suffix code
Option
code
Description
ZO21R
-L
- - - - - - Probe Protector(0 to 700 °C)
Insertion
length
-100
-150
-200
- - - - - -
- - - - - -
- - - - - -
1.05 m (3.5 ft)
1.55 m (5.1 ft)
2.05 m (6.8 ft)
Flange ( *1)
-J
-A
-E
- - - - - -
- - - - - -
- - - - - -
JIS 5K 65 FF SUS304
ANSI Class 150 4 FF SUS304
DIN PN10 DN50A
Style code
*B
- - - - - - Style B
*1 Thickness of flange depends on dimensions of flange.
F2-3E.ai
D
Ø
B
C
Ø
B
Flange <1>
(with bolts, nuts and washer)
gasket (t1.5)
SUS316
Dimensions of
holes on opposing surface
Ø
A
Gas flow
Washer (12)
Mounting nut (M12)
l (Insert length)
l=1050,1550,2050
Ø
60.5
Flange<1>
A
B
C
t
D
JIS 5K 65 FF SUS304
155
130
4 - Ø15
5
40
ANSI Class 150 4 FF SUS304
228.6
190.5
8 - Ø19
12
50
DIN PN10 DN50A SUS304
165
125
4 - Ш18
12
50