Using the sma trigger in on an avalanche, Using the bnc trigger out on sastracer – Teledyne LeCroy SAS Suite User Manual User Manual
Page 176
Version 6.25
SASSuite User Manual
166
LeCroy Corporation
Using the
SMA Trigger
In on an
Avalanche
Step 1 Attach an appropriate Trigger Out connector on a general
purpose test instrument to the SMA EXT IN connector using the
provided SMA cable or an SMA-to-BNC cable, as appropriate.
Step 2 In the SASSuite software, under Recording Options
(Avalanche)..., select the General Tab and click the radio
button Event Trigger under Recording Type.
Step 3 Select the Triggering/Filtering Tab. Under the Global Rules
section (or in one of the Sequence sections), click on the
Step 4 Select SMA Input, adjust the parameters to match the signal
expected from the external device, and click OK.
Step 5 Click on the
then click OK.
Step 6 Select Start from the Record menu.
Using the
BNC Trigger
Out on
SASTracer
Step 1 Attach the BNC Trigger Out connector to an appropriate BNC
Trigger In connector on a general purpose test instrument using
the provided BNC cable
Step 2 In the SASSuite software, under Recording Options (UPAS
10K)..., select the Recording Rules Tab
Step 3 Select the New Event button and specify the protocol level
event desired to initiate the external Trigger Out signal.