beautypg.com

21 timing violation analysis – Teledyne LeCroy Kibra DDR User Manual

Page 92

background image

Teledyne LeCroy

Timing Violation Analysis

90

 

 

Kibra DDR Protocol Analyzer User Manual

Figure 2.63: Row Usage Report Dialog

2.21

Timing Violation Analysis

The Timing Violation Analysis dialog (see 

Figure 2.64 on page 91

and 

Figure 2.65 on 

page 92

)displays the memory controller and timing violation parameters. Refer to 

“Recording Options ‐ Basic Settings Tab” on page 42

 for more information.

This allows the user to run the Violation Check on the trace data with revised values for 
the timing parameters. You can do "What if?" scenarios to see if the new values would 
have created violations had they been in place during the capture. The default for Follow 
MRS Settings
 is OFF in this feature, Please ensure it is turned on if you want the software 
to follow the MRS changes as it analyzes the trace. But in that case, be aware that the 
values you changed may be overridden by values obtained from the MRS commands.If 
the trace was captured with no filtering, a checkbox is provided to allow the user to 
change the CAL value.