Three samples are required:

Empty: Nothing in the sensor.

Gap (or Mark): The media liner, notch, or reflective mark in the sensor.

Paper: The label (and liner, if any) in the sensor.



With the correct SENSOR TYPE selected (see Section 4.2.2), perform a Standard Calibration

as follows:

Step

Action

Displayed Message

Comment

1

Turn ON the printer.

UNCALIBRATED

Wait briefly, about six

seconds, for the printer
to initialize.

2

Press the

Key, and

then raise the printhead

assembly.

MENU MODE

MEDIA SETTINGS

You are in MENU MODE.

3

Press the

Key.

MEDIA SETTINGS

MEDIA TYPE

You are in MEDIA

SETTINGS.

4

Press the

Key then

scroll to SENSOR
CALIBRATION.

MEDIA SETTINGS

SENSOR CALIBRATION

Press the

Key to

cancel this procedure.

5

Press the

Key.

SENSOR CALIBRATION

PERFORM CALIBRATION

You are beginning the

procedure.

6

Press the

Key, and

then press any key.

REMOVE LABEL STOCK

PRESS ANY KEY

No media should be in
the sensor. This sets the

Empty value, where
“yyy” is the current
sensor reading.