Fault Messages (continued)

Fault Messages

Displayed Message

Description

Action(s)

REFLECTIVE MODE
CANNOT CALIBRATE

Consistently low sensor
readings were detected.


Press any key to continue.


Ensure that the reflective mark is

inserted facedown in the media
sensor during calibration; also,
ensure that the reflective mark is

made of carbon based ink; and,
that the sensor is free of debris.

Retry calibration. If the problem
persists, try Advanced Entry

Calibration; see Section 5.2.2.

REFLECTIVE MODE
FAULTY SENSOR

Consistently high sensor

readings were detected.

Press any key to continue.

Ensure that media was removed

from the media sensor during the
appropriate calibration step; also

ensure that no labels are stuck in
the media sensor. Retry the

calibration. If the problem
persists, call for service.

RIBBON FAULT

One of the following has

occurred:

The ribbon sensor
values have changed;
or,

No ribbon supply hub
rotation or only

intermittent movement
has been detected.

Try the following procedures:

Ensure that ribbon is correctly
loaded and that the printhead
assembly is locked;

Remove any obstruction that
may inhibit ribbon hub

movement;

Ensure that the ribbon core
fits snugly on the ribbon
supply hub; and,

Ensure that the ribbon and
label combination is not
slipping; see Section 7.2.2 for

recommended combinations.