MONITOR UNIT STATUS

The following status signals are open collector TTL signals via an optocoupler referenced to
Programming Control port pin 8: maximum voltage is 30V, maximum current is 8mA. See Figure
3-7. Status noted in Figure 3-7 is present when signal is TTL high, absent when TTL low.

FIGURE 3-7. PROGRAMMING CONTROL PORT STATUS MONITORING

3.6

DIGITAL REMOTE MODE PROGRAMMING

KLN models may be digitally programmed over a control bus using SCPI (Standard Commands
for Programmable Instruments). SCPI provides a common language conforming to IEEE 488.2
for instruments used in an automatic test system (see PAR. 3.7). The control bus used must be
either the RS-485 Serial Bus, or the optional IEEE 488 standard communication bus (General
Purpose Interface Bus, GPIB) or the optional LAN interface. Refer to PAR. 3.7 for a detailed dis-
cussion of SCPI programming.

Refer to Table 2-3 for input/output signal allocations for communication via the GPIB and Table
2-2 for RS-485.

This section includes the requirements for

RS 485 Operation (see PAR. 3.6.1)

GPIB Operation if option installed (see PAR. 3.6.2) including an explanation of GPIB bus

protocols (PAR. 3.6.2.1)

LAN Operation if option installed (see PAR. 3.6.3)

3.6.1

RS-485 OPERATION

See PAR. 2.7.2 for RS-485 connections.

See PAR. 3.3.18 to change the RS-485 Address.

See PAR. 3.3.16 to change the RS-485 Baud Rate.

3.6.2

GPIB OPERATION (IF OPTION INSTALLED)

See PAR. 2.7.4 for GPIB connections.

See PAR. 3.3.17 to change the GPIB Address.

STATUS PIN

DESCRIPTION

7

Unit power on

9

Alarm (OVP, OCP or output shut-

down applied to pin 23)

10

Output on/off

11

Constant Current (CC)

12

Constant Voltage (CV)